www.pfeiffer-vacuum.net<br />

The following combination sensors are available:<br />

PCR 260 capacitive diaphragm-Pirani combination transmitter with a measurement<br />

range from 1,500 to 5 . 10 -4 mbar<br />

PKR 251 and 261 Pirani-cold cathode combination transmitters with measurement<br />

ranges from 1,000 to 5 . 10 -9 mbar<br />

PBR 260 Pirani/Bayard-Alpert combination transmitter with a measurement range<br />

from 1,000 to 5 . 10 -10 mbar.<br />

ActiveLine transmitters offer the following advantages:<br />

There are three control units for one, two or six transmitters<br />

Any transmitter can be operated with any control unit<br />

Uniform connection cable<br />

3.3.1.3 ModulLine<br />

The ModulLine series includes three TPR Pirani vacuum gauges with a measurement range<br />

from 1,000 to 8 . 10 - 4 mbar and three IKR cold cathode measuring tubes with measurement<br />

ranges from 5 . 10 - 3 to 10 -11 mbar. Because these vacuum gauges do not contain any electronics,<br />

they are suitable for use in high-radiation environments.<br />

TPG 300 Control Unit Up to three boards<br />

Figure 3.11: TPG 300 control unit for ModulLine vacuum gauges<br />

A TPG 300 with the following features serves as the control unit:<br />

Two measurement boards can be inserted, as well as<br />

An RS-232/RS-485 interface board with relay outputs<br />

Connection of up to four vacuum gauges<br />

Fieldbus connection is possible<br />

Measurement board<br />

Messkarte<br />

Measurement board<br />

Interface and relay<br />

board<br />

Page 85<br />

<strong>Vacuum</strong><br />

<strong>Technology</strong>