4.3.1.2 Dynamic chamber-air thermal balance subsystem<br />

The Dynamic chamber-air thermal balance subsystem consists of 3 subsystems. They<br />

are listed below:<br />

Fluid dynamic section subsystem (for dynamic fluctuating flow)<br />

Chamber air specific humidity subsystem<br />

Chamber-air fluctuating thermal balance subsystem<br />

The “Fluid dynamic section subsystem (for dynamic fluctuating flow)” is the same as<br />

the one in the Fluid dynamic and exhaled air re-breath model. The Chamber air specific<br />

humidity subsystem is the same as the one described in Appendix VIII. The differences<br />

are that the inputs are all fluctuating and its input of “specific humidity of inlet” is from<br />

a switch block. The block provides specific humidity of inlet from the blower when<br />

flow direction is from the blower and from the HADT lump 1 when the flow is reverse.<br />

For the Chamber-air fluctuating thermal balance subsystem, it is basically the same as<br />

the steady state chamber-air thermal balance subsystem described in Appendix VIII.<br />

The differences are also that the inputs are all fluctuating and the inlet specific humidity<br />

and inlet temperature are from the blower when flow is positive and from the HADT<br />

lump 1 when the flow is reverse.<br />

The inputs are listed in Table 4.17 below.<br />

Input port<br />

number<br />

Table 4.17 Inputs to the dynamic chamber-air thermal balance subsystem<br />

Input Unit<br />

1 Ambient air specific humidity In decimal<br />

2 Specific humidity of air from HADT due to reverse flow In decimal<br />

3 Temperature of inlet to the chamber from ADU °C<br />

4 Temperature of air from HADT due to reverse flow °C<br />

5 Breath load g/s<br />

6 Water temperature °C<br />

7 CPAP pressure setting cmH2O<br />

8 Portion of impact area to total water surface area In decimal<br />

9 Mask capacity m 3<br />

10 Temperature of wall 2 and 3 °C<br />

96