Table 5.1 shows the combinations of conditions and settings under which the<br />

experiment was conducted:<br />

Table 5.1 Combinations of conditions and settings for steady state thermal validation<br />

T (°C) Pressure<br />

Heating Element<br />

Setting<br />

Tube Heating Setting<br />

Low(about14°C) 4 cmH2O 45°C 0 W<br />

Normal(about22°C) 12 cmH2O 55°C 15 W<br />

High(about33°C) 20 cmH2O 65°C 30 W<br />

5.3.1.1 In-chamber water temperature<br />

After the whole system reached steady state and the ambient conditions were stable<br />

(when using the big environmental control room for high and low ambient temperature),<br />

water temperature was then measured by putting the thermal couple junction of the<br />

Fluke 51 II Thermometer into the water (Figure 5.15). The temperature was measured<br />

again after about 2.5 hours. The average of the two measurements was considered the<br />

result of the water temperature and was used to compare with the model output.<br />

Meanwhile, the ambient temperature and relative humidity were also measured twice.<br />

One was measured at the beginning of the test and the other at the end of the test. The<br />

experimental results and model outputs are shown in Figure 5.16, Figure 5.17 and<br />

Figure 5.18 with corresponding discussions.<br />

Figure 5.15 Measuring of airflow temperature at chamber outlet<br />

123