2.3.5 Heated Air Delivery Tube<br />

Figure 2.15 Corrugated wall of the HADT [19]<br />

The HADT is a corrugated tube 1.725 m in length. The corrugation of its wall is shown<br />

in Figure 2.15.<br />

The HADT and the mask fluid dynamic model are shown in Figure 2.16. The minor<br />

losses at the ends of the HADT are neglected for simplification. Since the density<br />

variation of air flow in the HADT is small, an average density of 1.142 kg/m 3 is<br />

assumed for simplification. The average density is based on the mean working<br />

conditions in the HADT as 30°C air temperature, 80% relative humidity and 12 cmH2O<br />

ADU pressure setting.<br />

Figure 2.16 controlled volume analysis of airflow in HADT<br />

24