Table III. 2 Airflow temperature and enthalpy increases after flowing through ADU when ambient<br />

temperature is 20°C and relative humidity is 70%<br />

P-set(cmH2O) T ( C)<br />

Ao<br />

- 185 -<br />

T ( C)<br />

Ao<br />

H<br />

( J / kg)<br />

4 22.89 2.89 2928.67<br />

6 22.93 2.93 2969.04<br />

8 23.20 3.20 3239.90<br />

10 23.73 3.73 3780.83<br />

12 24.40 4.40 4460.65<br />

14 25.19 5.19 5262.07<br />

16 25.92 5.92 6002.51<br />

18 26.66 6.66 6746.34<br />

20 27.03 7.03 7129.71<br />

When under temperature 35°C and relative humidity 90%, the specific humidity level is<br />

d = 0.03167 (kg/kg) and the result is shown below:<br />

Table III. 3 Airflow temperature and enthalpy increases after flowing through ADU when ambient<br />

temperature is 35°C and relative humidity is 90%<br />

P-set(cmH2O) T ( C)<br />

Ao<br />

T ( C)<br />

Ao<br />

H<br />

( J / kg)<br />

4 37.89 2.89 2978.92<br />

6 38.46 3.46 3568.89<br />

8 39.01 4.01 4140.59<br />

10 39.58 4.58 4727.30<br />

12 40.15 5.15 5310.94<br />

14 40.66 5.66 5836.77<br />

16 41.00 6.00 6197.36<br />

18 41.37 6.37 6578.09<br />

20 41.44 6.44 6642.78