Figure 6.1 Reverse flow under different combinations of breath load and CPAP pressure setting<br />

6.2.1.2 Mask capacity vs. reverse flow<br />

The mask capacity in the model is adjustable. However, the model output shows that the<br />

distance of reverse flow is the same when mask capacity changes. This is because of the<br />

assumption of air incompressibility which means the distance of reverse flow is not<br />

influenced by the mask capacity.<br />

6.2.2 Exhaled air re-inhalation<br />

In this section, the breath load, pressure setting, the resulted reverse flow and mask<br />

capacity are discussed with respect to their influences on the exhaled air re-breathing.<br />

6.2.2.1 Exhaled air re-inhalation vs. breath load and pressure setting<br />

The percentage of exhaled air in inhalation vs. combination of breath load and pressure<br />

settings using the nasal mask is shown in Figure 6.2.<br />

147