Figure I. 1 Tested results and the regressed curve for ADU outlet pressure when pressure setting is<br />

4cmH2O<br />

By using the same method, such a 3-ordered expression can be obtained for each of the<br />

tested pressure settings. The coefficients for each pressure setting are listed below:<br />

P Ao a b c d<br />

4cmH2O -1.1387 3.2879 -10.0609 415.40<br />

6cmH2O -1.1219 3.8749 -9.4252 611.44<br />

8cmH2O -1.3316 5.8853 -10.4933 807.36<br />

10cmH2O -1.4676 7.2633 -9.4144 1001.50<br />

12cmH2O -1.6729 9.3706 -11.9589 1187.60<br />

14cmH2O -1.8456 10.7920 -10.7130 1378.05<br />

16cmH2O -2.1916 14.0150 -14.9430 1563.70<br />

18cmH2O -2.4434 15.5360 -9.6832 1748.35<br />

20cmH2O -2.5066 16.9690 -12.4421 1957.05<br />

It is convenient for dynamic simulation modelling to unify them by regressing the<br />

coefficients again as:<br />

a 0.0961P 0.5932<br />

i set<br />

b 0.9124P 1.2827<br />

i set<br />

c 0.1791P <br />

8.9763<br />

i set<br />