Figure 5.30 Comparison of model outputs and experimental results of Airflow temperature at<br />

HADT lump30 vs. setting of tube heating under heating element temperature setting at 55°C,<br />

pressure setting at 12cmH2O and normal room temperature<br />

From the data and graphs above, the ambient temperature and tube heating are the most<br />

important factors influencing the airflow temperature at the HADT outlet.<br />

The model outputs match very well with the experimental results.<br />

5.3.1.5 Condensation in the HADT<br />

Condensation in the HADT was checked by observation. The photo below is used to<br />

define the terms describing the condensation severity (Figure 5.31).<br />