8. EXPLOSION PROTECTED TYPE INSTRUMENT<br />

Maximum surface temperature:<br />

Maximum Surface<br />

Temperature<br />

T75°C (+167°F)<br />

T85°C (+185°F)<br />

T100°C (+212°F)<br />

T110°C (+230°F)<br />

Maximum Process<br />

Temperature<br />

+70°C (+158°F)<br />

+85°C (+185°F)<br />

+120°C (+248°F)<br />

+130°C (+266°F)<br />

T0802.EPS<br />

Ambient Temp.:–40°C to +60°C (–40°F to +140°F)<br />

(Remote Flowtube)<br />

CENELEC ATEX (KEMA) Flameproof Type<br />

Group: II<br />

Category: 2G<br />

EEx dme [ia] IIC T6...T3<br />

Electrode Circuit Um: 250 Vac/dc<br />

Excitation Circuit: 170V max<br />

Enclosure: IP66, IP67<br />

Temperature Class:<br />

(2) Electrical Connection<br />

The type of electrical connection is stamped near the<br />

electrical connection port according to the following<br />

codes.<br />

(<strong>Integral</strong> <strong>Flowmeter</strong>)<br />

Screw Size<br />

ISO M20x1.5 female<br />

ANSI 1/2NPT female<br />

(Remote Flowtube)<br />

Screw Size<br />

ISO M20x1.5 female<br />

ANSI 1/2NPT female<br />

Marking<br />

M<br />

A<br />

Marking<br />

M<br />

A<br />

Temperature<br />

Class<br />

Maximum Process<br />

Temperature<br />

Minimum Process<br />

Temperature<br />

T6<br />

T5<br />

T4<br />

T3<br />

+70°C (+158°F)<br />

+85°C (+185°F)<br />

+120°C (+248°F)<br />

+150°C (+302°F)<br />

–40°C (–40°F)<br />

–40°C (–40°F)<br />

–40°C (–40°F)<br />

–40°C (–40°F)<br />

T0803.EPS<br />

Ambient Temp.: –40°C to +60°C (–40°F to +140°F)<br />

CENELEC ATEX (KEMA) Type of Protection<br />

“Dust”<br />

Group: II<br />

Category: 1D<br />

Electrode Circuit Um: 250 Vac/dc<br />

Excitation Circuit: 170V max<br />

Enclosure: IP66, IP67<br />

Maximum surface temperature:<br />

(3) Installation<br />

WARNING<br />

F0801.EPS<br />

•All wiring shall comply with local installation<br />

requirements and local electrical code.<br />

• In hazardous locations, the cable entry devices<br />

shall be of a certified ATEX flameproof type,<br />

suitable for the conditions of use and correctly<br />

installed.<br />

• Unused apertures shall be closed with suitable<br />

flameproof certified blanking elements. (The<br />

plug attached is flameproof certified.)<br />

Maximum Surface<br />

Temperature<br />

T75°C (+167°F)<br />

T85°C (+185°F)<br />

T100°C (+212°F)<br />

T115°C (+239°F)<br />

Maximum Process<br />

Temperature<br />

+70°C (+158°F)<br />

+85°C (+185°F)<br />

+120°C (+248°F)<br />

+150°C (+302°F)<br />

T0804.EPS<br />

Ambient Temp.:–40°C to +60°C (–40°F to +140°F)<br />

(4) Operation<br />

(<strong>Integral</strong> <strong>Flowmeter</strong>)<br />

WARNING<br />

• After de-energizing, delay 20 minutes before<br />

opening.<br />

• Take care not to generate mechanical spark<br />

when access to the instrument and peripheral<br />

devices in hazardous locations.<br />

8-2<br />

IM 01E20D01-01E