5. MAINTENANCE<br />
(6) Move aside the electrode lead wire and avoiding<br />
the screw, insert end A of the tool to the electrode<br />
and loosen the electrode holder using the torque<br />
driver.<br />
5.2.2 Cleaning Replaceable Electrodes<br />
(1) Clean the electrode surface (wetted part) with<br />
alcohol or other cleaning agents.<br />
O-ring<br />
Electrode surface<br />
Electrode holde<br />
End A<br />
Torque driver<br />
F0506.EPS<br />
(7) Screw end B of the tool into the screw block of the<br />
electrode, and pull the tool straight out to remove<br />
the electrode holder.<br />
F0508.EPS<br />
NOTE<br />
The screw threads are coated with an antiseizing<br />
compound. Do not let the compound<br />
come in contact with the electrode.<br />
(2) Clean the surface of the lining seal with an alcohol<br />
moistened swab.<br />
End B<br />
F0507.EPS<br />
Lining seal surface<br />
F0509.EPS<br />
5-3<br />
IM 01E20D01-01E