5. MAINTENANCE<br />
5. MAINTENANCE<br />
WARNING<br />
• Maintenance work must be carried out by<br />
expert engineer or skilled personnel and not by<br />
operators.<br />
• Before opening the cover, it is important to<br />
ensure that at least 10 minutes have passed<br />
since the power was turned off. Furthermore,<br />
opening of the cover must also be carried out<br />
by expert engineer or skilled personnel.<br />
CAUTION<br />
• Explosion protected type must be, as a rule,<br />
removed to a non-hazardous area for maintenance<br />
and be disassembled and reassembled<br />
to the original state.<br />
• The cover is locked by the special screw. In<br />
case of opening the cover, use the hexagonal<br />
wrench attached.<br />
• Be sure to lock the cover by the special screw<br />
using the hexagonal wrench attached after<br />
installing the cover.<br />
5.1 Changing Direction of<br />
Electrical Connection<br />
F0501.EPS<br />
(4) Using the hexagonal wrench, loosen the screw in<br />
the neck.<br />
(5) Turn the converter or the terminal box in the<br />
desired direction.<br />
NOTE<br />
The converter and the terminal box can be<br />
turned –140 degree to +180 degree from the<br />
arrow mark indicating the flow direction. Do not<br />
exceed these angle.<br />
(6) Using the hexagonal wrench, retighten the neck<br />
screw.<br />
IMPORTANT<br />
The following types can not be changed direction<br />
of electrical connection after delivery.<br />
• Submersible Type.<br />
• Optional code DHC (for district heating and<br />
cooling or condensation-proof).<br />
• Size of 1100 mm (44 in.) or larger.<br />
(1) The following tools are required to change the<br />
direction of the electrical connection:<br />
• Hexagonal wrench (nominal size 1.5): Comes<br />
with the instrument.<br />
• Wrench<br />
(2) Turn off the power to the flowmeter.<br />
(3) Using the wrench, loosen the hexagonal nut at the<br />
neck of the instrument.<br />
F0502.EPS<br />
(7) Using the wrench, retighten the hexagonal nut at<br />
the neck. After that, check that the converter or<br />
terminal box is fixed.<br />
5-1<br />
IM 01E20D01-01E