User's Manual AXF Magnetic Flowmeter Integral ... - Yokogawa
User's Manual AXF Magnetic Flowmeter Integral ... - Yokogawa
User's Manual AXF Magnetic Flowmeter Integral ... - Yokogawa
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
8. EXPLOSION PROTECTED TYPE INSTRUMENT<br />
(3) Operation<br />
(<strong>Integral</strong> <strong>Flowmeter</strong>)<br />
WARNING<br />
• After de-energizing, delay 20 minutes before<br />
opening.<br />
• Take care not to generate mechanical spark<br />
when access to the instrument and peripheral<br />
devices in hazardous locations.<br />
(Remote Flowtube)<br />
WARNING<br />
• De-energize before opening.<br />
• Take care not to generate mechanical spark<br />
when access to the instrument and peripheral<br />
devices in hazardous locations.<br />
(4) Maintenance and Repair<br />
WARNING<br />
The instrument modification or parts replacement<br />
by other than authorized representative of<br />
<strong>Yokogawa</strong> Electric Corporation is prohibited and<br />
will void the certification.<br />
8.5 TIIS<br />
CAUTION<br />
The model <strong>AXF</strong>C magnetic flowmeter with<br />
optional code JF3, which has obtained<br />
certification according to technical criteria for<br />
explosion-protected construction of electric<br />
machinery and equipment (Standards<br />
Notification No. 556 from the Japanese Ministry<br />
of Labor) conforming to IEC standards, is<br />
designed for hazardous areas where<br />
inflammable gases or vapors may be present.<br />
(This allows installation in Division 1 and 2<br />
areas)<br />
To preserve the safety of flameproof equipment<br />
requires great care during mounting, wiring, and<br />
piping. Safety requirements also place<br />
restrictions on maintenance and repair activities.<br />
Users absolutely must read “INSTALLATION<br />
AND OPERATING PRECAUTIONS FOR TIIS<br />
FLAMEPROOF EQUIPMENT” at the end of this<br />
manual.<br />
(1) Technical Data<br />
Certificate:<br />
Size: mm<br />
(inch)<br />
Lining<br />
2.5 (0.1)<br />
5 (0.2)<br />
10 (0.4)<br />
15 (0.5)<br />
25 (1.0)<br />
32 (1.25)<br />
40 (1.5)<br />
50 (2.0)<br />
65 (2.5)<br />
80 (3.0)<br />
100 (4.0)<br />
125 (5.0)<br />
150 (6.0)<br />
200 (8.0)<br />
250 (10)<br />
300 (12)<br />
350 (14)<br />
400 (16)<br />
<strong>Integral</strong> <strong>Flowmeter</strong><br />
PFA<br />
Lining<br />
C16630<br />
C16630<br />
C16630<br />
C16630<br />
C16631<br />
C16632<br />
C16633<br />
C16634<br />
C16635<br />
C16636<br />
C16637<br />
C16638<br />
C16639<br />
C16640<br />
C16641<br />
C16642<br />
C16643<br />
C16644<br />
Ceramics<br />
Lining<br />
C16645<br />
C16645<br />
C16645<br />
C16646<br />
C16647<br />
—<br />
C16648<br />
C16649<br />
—<br />
C16650<br />
C16651<br />
—<br />
C16652<br />
C16653<br />
—<br />
—<br />
—<br />
—<br />
Remote Flowtube<br />
PFA<br />
Lining<br />
C16654<br />
C16654<br />
C16654<br />
C16654<br />
C16655<br />
C16656<br />
C16657<br />
C16658<br />
C16659<br />
C16660<br />
C16661<br />
C16662<br />
C16663<br />
C16664<br />
C16665<br />
C16666<br />
C16667<br />
C16668<br />
Ceramics<br />
Lining<br />
C16669<br />
C16669<br />
C16669<br />
C16670<br />
C16671<br />
—<br />
C16672<br />
C16673<br />
—<br />
C16674<br />
C16675<br />
—<br />
C16676<br />
C16677<br />
—<br />
—<br />
—<br />
—<br />
T33.EPS<br />
(<strong>Integral</strong> <strong>Flowmeter</strong>)<br />
• Construction: Ex de[ia] IIC T4<br />
: Converter ; Explosion proof<br />
Flowtube ; Increased Safety and<br />
Intrinsically Safety(ia)<br />
Electrode ; Intrinsically Safety(ia)<br />
8-8<br />
IM 01E20D01-01E