8. EXPLOSION PROTECTED TYPE INSTRUMENT<br />

(3) Operation<br />

(<strong>Integral</strong> <strong>Flowmeter</strong>)<br />

WARNING<br />

• After de-energizing, delay 20 minutes before<br />

opening.<br />

• Take care not to generate mechanical spark<br />

when access to the instrument and peripheral<br />

devices in hazardous locations.<br />

(Remote Flowtube)<br />

WARNING<br />

• De-energize before opening.<br />

• Take care not to generate mechanical spark<br />

when access to the instrument and peripheral<br />

devices in hazardous locations.<br />

(4) Maintenance and Repair<br />

WARNING<br />

The instrument modification or parts replacement<br />

by other than authorized representative of<br />

<strong>Yokogawa</strong> Electric Corporation is prohibited and<br />

will void the certification.<br />

8.5 TIIS<br />

CAUTION<br />

The model <strong>AXF</strong>C magnetic flowmeter with<br />

optional code JF3, which has obtained<br />

certification according to technical criteria for<br />

explosion-protected construction of electric<br />

machinery and equipment (Standards<br />

Notification No. 556 from the Japanese Ministry<br />

of Labor) conforming to IEC standards, is<br />

designed for hazardous areas where<br />

inflammable gases or vapors may be present.<br />

(This allows installation in Division 1 and 2<br />

areas)<br />

To preserve the safety of flameproof equipment<br />

requires great care during mounting, wiring, and<br />

piping. Safety requirements also place<br />

restrictions on maintenance and repair activities.<br />

Users absolutely must read “INSTALLATION<br />

AND OPERATING PRECAUTIONS FOR TIIS<br />

FLAMEPROOF EQUIPMENT” at the end of this<br />

manual.<br />

(1) Technical Data<br />

Certificate:<br />

Size: mm<br />

(inch)<br />

Lining<br />

2.5 (0.1)<br />

5 (0.2)<br />

10 (0.4)<br />

15 (0.5)<br />

25 (1.0)<br />

32 (1.25)<br />

40 (1.5)<br />

50 (2.0)<br />

65 (2.5)<br />

80 (3.0)<br />

100 (4.0)<br />

125 (5.0)<br />

150 (6.0)<br />

200 (8.0)<br />

250 (10)<br />

300 (12)<br />

350 (14)<br />

400 (16)<br />

<strong>Integral</strong> <strong>Flowmeter</strong><br />

PFA<br />

Lining<br />

C16630<br />

C16630<br />

C16630<br />

C16630<br />

C16631<br />

C16632<br />

C16633<br />

C16634<br />

C16635<br />

C16636<br />

C16637<br />

C16638<br />

C16639<br />

C16640<br />

C16641<br />

C16642<br />

C16643<br />

C16644<br />

Ceramics<br />

Lining<br />

C16645<br />

C16645<br />

C16645<br />

C16646<br />

C16647<br />

—<br />

C16648<br />

C16649<br />

—<br />

C16650<br />

C16651<br />

—<br />

C16652<br />

C16653<br />

—<br />

—<br />

—<br />

—<br />

Remote Flowtube<br />

PFA<br />

Lining<br />

C16654<br />

C16654<br />

C16654<br />

C16654<br />

C16655<br />

C16656<br />

C16657<br />

C16658<br />

C16659<br />

C16660<br />

C16661<br />

C16662<br />

C16663<br />

C16664<br />

C16665<br />

C16666<br />

C16667<br />

C16668<br />

Ceramics<br />

Lining<br />

C16669<br />

C16669<br />

C16669<br />

C16670<br />

C16671<br />

—<br />

C16672<br />

C16673<br />

—<br />

C16674<br />

C16675<br />

—<br />

C16676<br />

C16677<br />

—<br />

—<br />

—<br />

—<br />

T33.EPS<br />

(<strong>Integral</strong> <strong>Flowmeter</strong>)<br />

• Construction: Ex de[ia] IIC T4<br />

: Converter ; Explosion proof<br />

Flowtube ; Increased Safety and<br />

Intrinsically Safety(ia)<br />

Electrode ; Intrinsically Safety(ia)<br />

8-8<br />

IM 01E20D01-01E