6. OUTLINE<br />
■ ACCESSORIES<br />
Remote Flowtube (size 2.5 to 1000 mm(0.1 to 40 in.)):<br />
Centering device (wafer type only): 1 pc.<br />
Hexagonal wrench: 2 pcs.<br />
<strong>Integral</strong> <strong>Flowmeter</strong>:<br />
Centering device (wafer type only): 1 pc.<br />
Fuse (T2.0A, 250 V): 1 pc.<br />
*Time lag fuse<br />
Hexagonal wrench: 2 pcs.<br />
Remote Flowtube<br />
Terminal configuration<br />
EX1 EX2<br />
A B C<br />
size 2.5 to 400 mm (0.1 to 16 in.)<br />
EX1 EX2<br />
A B C<br />
(Only for Explosion proof type)<br />
■ TERMINAL CONFIGURATION AND<br />
TERMINAL WIRING<br />
C<br />
EX1<br />
C<br />
<strong>Integral</strong> <strong>Flowmeter</strong> “”<br />
EX1<br />
EX2<br />
B<br />
EX2<br />
A<br />
A<br />
B<br />
Terminal configuration<br />
size 500 to 1000 mm (20 to 40 in.)<br />
size 1100 to 2600 mm (44 to 104 in.)<br />
Terminal<br />
Symbols<br />
N/ -<br />
L/ +<br />
I +<br />
I -<br />
DO +<br />
DO -<br />
DIO +<br />
DIO -<br />
POWER<br />
SUPPLY<br />
DO<br />
DIO<br />
+ L/<br />
N/<br />
- +<br />
+<br />
+<br />
Terminal wiring<br />
Description<br />
Functional grounding<br />
Power supply<br />
Current output 4 to 20mA DC<br />
Pulse output/Alarm output/<br />
Status output<br />
Alarm output/Status output<br />
Status input<br />
Protective grounding<br />
(Outside of the terminal)<br />
F41.EPS<br />
Terminal wiring<br />
Terminal<br />
Symbols<br />
A<br />
B<br />
C<br />
EX1<br />
EX2<br />
Description<br />
Flow signal output<br />
Excitation current input<br />
Functional grounding<br />
(Only for explosion proof type)<br />
Protective grounding<br />
(Outside of the terminal)<br />
Note: When submersible type or option code DHC is<br />
selected, waterproof glands and a 30m long cable<br />
are attached.<br />
F42.EPS<br />
6-20<br />
IM 01E20D01-01E