3. INSTALLATION<br />
Table 3.3.3<br />
Tightening torque values for Union Joint<br />
Type and Permeable Fluids<br />
Gasket<br />
Connecting fitting<br />
Union joint nut<br />
*Piping<br />
Size mm (inch) Torque (N-m / {kgf-cm} / [in-lbf])<br />
2.5 (0.1) 11 to 15 / {112.2 to 153} / [97.36 to 132.8]<br />
5 (0.2) 11 to 15 / {112.2 to 153} / [97.36 to 132.8]<br />
10 (0.4) 17 to 23 / {173.4 to 234.5} / [150.5 to 203.6]<br />
T0304.EPS<br />
3.3.2 Nominal Diameter 2.5 mm (0.1<br />
in.) to 40 mm (1.5 in.), Wafer<br />
Type<br />
*Mounting base<br />
*: To be provided by user<br />
F0309a.EPS<br />
IMPORTANT<br />
Use bolts and nuts in compliance with the flange<br />
ratings. When stud-type through-bolts are used,<br />
be sure the outside diameter of the shank is<br />
smaller than that of the thread ridge. Be sure to<br />
choose a gasket with an inner diameter that<br />
does not protrude inside the piping (refer to<br />
Table 3.3.13). If the inner diameter of the gasket<br />
is too large, however, fluid leakage may result.<br />
Horizontal mounting<br />
F0309b.EPS<br />
(1) Mounting Direction<br />
Mount the flowmeter so that the flow direction of the<br />
fluid to be measured is in line with the direction of the<br />
arrow mark on the flowmeter.<br />
IMPORTANT<br />
Figure 3.3.1<br />
Vertical mounting<br />
F0309c.EPS<br />
Mounting Procedure for Union Joint Type<br />
Table 3.3.2 Tightening torque values for Union Joint<br />
Type<br />
Apply these tightening torque values when the gaskets<br />
are Valqua #7020 (standard) or alkali-resistant gaskets<br />
for the metal piping (optional code GF).<br />
Size mm (inch) Torque (N-m / {kgf-cm} / [in-lbf])<br />
2.5 (0.1) 9 to 12 / {91.77 to 122.4} / [79.66 to 106.2]<br />
5 (0.2) 9 to 12 / {91.77 to 122.4} / [79.66 to 106.2]<br />
10 (0.4) 14 to 18 / {142.8 to 183.5} / [123.9 to 159.3]<br />
T0303.EPS<br />
If it is impossible to match the direction of the<br />
arrow mark, the direction of the electrical connection<br />
can be changed. Refer to Section 5.1 to<br />
do this properly.<br />
In case the fluid being measured flows against<br />
the arrow direction, refer to the parameter J20:<br />
Flow Direction in the user’s manual of the<br />
<strong>AXF</strong>A11 <strong>Magnetic</strong> <strong>Flowmeter</strong> Remote Converter<br />
(IM 01E20C01-01E) or the <strong>AXF</strong>A14 <strong>Magnetic</strong><br />
<strong>Flowmeter</strong> Remote Converter/<strong>AXF</strong> <strong>Integral</strong><br />
<strong>Flowmeter</strong> [Software Edition] (IM 01E20C02-<br />
01E).<br />
(2) Mounting Centering Devices<br />
To maintain concentricity of the flowmeter with the<br />
pipes, install centering devices on the Mini-flanges of<br />
the flowmeter. Use the appropriate centering devices<br />
according to the nominal diameter and the flange<br />
ratings.<br />