5.8 Troubleshooting<br />

5.8.1 No Indication<br />

5. MAINTENANCE<br />

Although magnetic flowmeters rarely require maintenance, failures may occur when the<br />

instrument is not operated correctly. This section describes troubleshooting procedures<br />

where the cause of the breakdown is identified through receiver indication.<br />

START<br />

Is an error being displayed? Check the converter display.<br />

Is an error being<br />

displayed?<br />

NO<br />

YES<br />

Refer to the “Alarm Functions” in<br />

the <strong>AXF</strong>A11, <strong>AXF</strong>A14, and <strong>AXF</strong><br />

<strong>Integral</strong> <strong>Flowmeter</strong> [Software<br />

Edition] user manuals.<br />

Is anything else<br />

being displayed?<br />

NO<br />

YES<br />

Is the fuse blown?<br />

NO<br />

YES<br />

Does the<br />

display return when the fuse<br />

is replaced?<br />

YES<br />

NO<br />

Does the<br />

arrow on the tube match the<br />

flow direction?<br />

YES<br />

Check whether the flowtube<br />

is filled with fluid.<br />

NO<br />

End<br />

Perform either of the following:<br />

a. Reinstall the flowtube in<br />

order to make the arrow and<br />

flow direction match.<br />

b. Reverse the setting for the<br />

flow direction.<br />

Is the flowtube<br />

full of fluid?<br />

NO<br />

YES<br />

Check the resistance between<br />

the flowtube’s A, B, and C<br />

terminals using a multimeter.* 1<br />

Is the<br />

resistance 100 M or more<br />

in each case?<br />

NO<br />

Change the piping or the<br />

mounting condition so that<br />

the tube fills with the fluid.<br />

*1: Refer to Section 5.7 for how to check<br />

the resistance of the remote type<br />

flowtube. The integral flowmeter's<br />

resistance cannot be checked.<br />

YES<br />

Remove the tube and<br />

inspect the inside of the pipe.<br />

Is foreign<br />

matter adhering to<br />

the walls?<br />

NO<br />

YES<br />

Remove the foreign matter.<br />

Contact a service center.<br />

F0521.EPS<br />

5-12<br />

IM 01E20D01-01E