5.8 Troubleshooting<br />
5.8.1 No Indication<br />
5. MAINTENANCE<br />
Although magnetic flowmeters rarely require maintenance, failures may occur when the<br />
instrument is not operated correctly. This section describes troubleshooting procedures<br />
where the cause of the breakdown is identified through receiver indication.<br />
START<br />
Is an error being displayed? Check the converter display.<br />
Is an error being<br />
displayed?<br />
NO<br />
YES<br />
Refer to the “Alarm Functions” in<br />
the <strong>AXF</strong>A11, <strong>AXF</strong>A14, and <strong>AXF</strong><br />
<strong>Integral</strong> <strong>Flowmeter</strong> [Software<br />
Edition] user manuals.<br />
Is anything else<br />
being displayed?<br />
NO<br />
YES<br />
Is the fuse blown?<br />
NO<br />
YES<br />
Does the<br />
display return when the fuse<br />
is replaced?<br />
YES<br />
NO<br />
Does the<br />
arrow on the tube match the<br />
flow direction?<br />
YES<br />
Check whether the flowtube<br />
is filled with fluid.<br />
NO<br />
End<br />
Perform either of the following:<br />
a. Reinstall the flowtube in<br />
order to make the arrow and<br />
flow direction match.<br />
b. Reverse the setting for the<br />
flow direction.<br />
Is the flowtube<br />
full of fluid?<br />
NO<br />
YES<br />
Check the resistance between<br />
the flowtube’s A, B, and C<br />
terminals using a multimeter.* 1<br />
Is the<br />
resistance 100 M or more<br />
in each case?<br />
NO<br />
Change the piping or the<br />
mounting condition so that<br />
the tube fills with the fluid.<br />
*1: Refer to Section 5.7 for how to check<br />
the resistance of the remote type<br />
flowtube. The integral flowmeter's<br />
resistance cannot be checked.<br />
YES<br />
Remove the tube and<br />
inspect the inside of the pipe.<br />
Is foreign<br />
matter adhering to<br />
the walls?<br />
NO<br />
YES<br />
Remove the foreign matter.<br />
Contact a service center.<br />
F0521.EPS<br />
5-12<br />
IM 01E20D01-01E