5. MAINTENANCE<br />
5.2 Removing, Cleaning,<br />
and Installing Replaceable<br />
Electrodes<br />
(General-Purpose<br />
Use Type Only)<br />
5.2.1 Removing Replaceable Electrodes<br />
(1) The following tools are required to replace the<br />
electrodes:<br />
• Special tool for removing and installing electrodes<br />
(F9807SK): Optional<br />
• Torque wrench or torque driver (nominal size 12)<br />
• Phillips screwdriver<br />
(4) Using end A of the tool, open the electrode cover.<br />
End A<br />
End B<br />
F0504.EPS<br />
(5) Hold down the electrode lead wire and terminal<br />
lug and remove the mounting screw using the<br />
Phillips screwdriver.<br />
F9807SK<br />
F0503.EPS<br />
(2) Turn off the power to the flowmeter.<br />
(3) Drain the fluid from the interior of the flowtube.<br />
Electrode lead wire:<br />
Red: Electrode A<br />
White: Electrode B<br />
WARNING<br />
• Never fail to drain the fluid from the interior of<br />
the flowtube.<br />
• If the fluid is at a high temperature, the instrument<br />
itself may also be hot. Be careful not to<br />
get burned when removing the electrodes.<br />
• When the process fluid is hazardous to humans,<br />
exercise caution to avoid coming into<br />
contact with it, and avoid inhaling any residual<br />
gas.<br />
NOTE<br />
F0505.EPS<br />
Always hold down the lead wire and terminal lug<br />
when removing the mounting screw.<br />
5-2<br />
IM 01E20D01-01E