5. MAINTENANCE<br />

5.2.3 Installing Replaceable Electrodes<br />

(2) Using end A of the tool, tighten the electrode<br />

holder.<br />

NOTE<br />

It is recommended to replace the O-ring when<br />

reinstalling the replaceable electrode. Use the O-<br />

ring specified by <strong>Yokogawa</strong> (G9303SE: material<br />

is fluororubber).<br />

NOTE<br />

Precautions for storage of O-rings:<br />

• Keep them in a cool, dark place.<br />

• Wrap them well.<br />

• Do not use O-rings after one year since their<br />

purchase.<br />

Torque driver<br />

NOTE<br />

F0511.EPS<br />

(1) Push the electrode holder with end B of the tool<br />

screwed into it, straight into the flowtube.<br />

Electrode lead wire:<br />

Red: Electrode A<br />

White: Electrode B<br />

Electrode holder:<br />

(Electrode A is engraved with "A"<br />

and electrode B has no markings.)<br />

Using the torque wrench or torque driver, tighten<br />

the electrode holder to the following torque<br />

values:<br />

PFA lining:<br />

8.0 N·m ±1 N·m (80 kgf·cm ±10 kgf·cm)<br />

Polyurethane rubber lining:<br />

3.0 N·m ±0.5 N·m (30 kgf·cm ±5 kgf·cm)<br />

WARNING<br />

If the electrode holder was not tightened to its<br />

specified torque value, fluid leakage from the<br />

electrode may result. Should fluid spill over the<br />

electrode block, wipe it dry and check that the<br />

O-ring has not come off.<br />

F0510.EPS<br />

NOTE<br />

• Check that the O-ring has not come off.<br />

• Do not interchange electrodes A and B when<br />

installing them. Install electrode A to the<br />

electrode boss with the red lead wire and<br />

electrode B to the one with the white lead wire.<br />

(3) Fill the interior of the flowtube with the fluid and<br />

check that there is no leakage from the electrode.<br />

(4) Hold down the lead wire and terminal lug and<br />

screw the mounting screw into the electrode using<br />

the Phillips screwdriver.<br />

Electrode mounting screw<br />

F0512.EPS<br />

5-4<br />

IM 01E20D01-01E