6. OUTLINE<br />
Natural Soft Rubber Lining<br />
General-Purpose Use and Submersible Type, Remote<br />
Flowtube (electrode structure code 1: Non-replaceable<br />
electrode)<br />
Pressure<br />
MPa (psi)<br />
4(580)<br />
2(290)<br />
1(145)<br />
50 mm (2.0 in.) (flange type, wafer type)<br />
65 to 200 mm (2.5 to 8.0 in.) (flange type, wafer type)<br />
250, 300 mm (10, 12 in.) (flange type)<br />
250, 300 mm (10, 12 in.) (wafer type)<br />
350, 400 mm (14 , 16 in.) (flange type)<br />
*1<br />
– 0.05(–7.3)<br />
–10 0 40 80<br />
(14) (32) (104) (175)<br />
Temperature °C (°F)<br />
*1 : –0.04 MPa (–5.7 psi) for sizes of 350 mm (14 in.) and 400 mm (16 in.)<br />
F05-2.EPS<br />
EPDM Rubber Lining<br />
General-Purpose Use and Submersible Type, Remote<br />
Flowtube (electrode structure code 1: Non-replaceable<br />
electrode)<br />
Pressure<br />
MPa (psi)<br />
4(580)<br />
2(290)<br />
1(145)<br />
50 mm (2.0 in.) (flange type, wafer type)<br />
65 to 200 mm (2.5 to 8.0 in.) (flange type, wafer type)<br />
250, 300 mm (10, 12 in.) (flange type)<br />
250, 300 mm (10, 12 in.) (wafer type)<br />
350, 400 mm (14, 16 in.) (flange type)<br />
*1<br />
– 0.05(–7.3)<br />
–10 0 40 80<br />
(14) (32) (104) (175)<br />
Temperature °C (°F)<br />
*1 : –0.04 MPa (–5.7 psi) for sizes of 350 mm (14 in.) and 400 mm (16 in.)<br />
F05-3.EPS<br />
Reasonable Figure for Thermal Shock of Creamics Lining:<br />
Size 2.5 to 25 mm (0.1 to 1.0 in.)<br />
∆T°C(°F)<br />
+120 (+248)<br />
+100 (+212)<br />
+60 (+140)<br />
Increase<br />
Decrease<br />
1(3) 5(15) 10(30)<br />
Flow Velocity m/s(ft/s)<br />
Size 40 and 50 mm (1.5 and 2.0 in.)<br />
∆T°C(°F)<br />
Increase<br />
+100 (+212)<br />
+90 (+162)<br />
+50 (+122)<br />
+30 (+86)<br />
Decrease<br />
1(3) 5(15) 10(30)<br />
Flow Velocity m/s(ft/s)<br />
Size 80 to 200 mm (3.0 to 8.0 in.)<br />
∆T°C(°F)<br />
Increase<br />
+70 (+158)<br />
+50 (+122)<br />
+30 (+86)<br />
Decrease<br />
1(3) 5(15) 10(30)<br />
Flow Velocity m/s(ft/s)<br />
F21.EPS<br />
“Decrease” means that the temperature of a measured<br />
fluid drops rapidly, while “increase” means that the<br />
temperature rises rapidly. The maximum allowable ranges<br />
in both cases are indicated by the curves shown in the<br />
diagrams, with the solid line indicating the maximum<br />
increase, and the broken line the maximum decrease.<br />
∆T: Change in temperature of measured fluid in one<br />
second<br />
Flow velocity: flow velocity of the measured fluid<br />
Allowable Conditions for Cleaning Sanitary Type<br />
Linings<br />
Steam or hot water cleaning: Max.temp.= +150 °C<br />
(+302°F), time= 60 minutes or less<br />
Vibration Conditions:<br />
Level of vibration in conformity with IEC 60068-2-6<br />
(SAMA 31.1-1980)<br />
• <strong>Integral</strong> <strong>Flowmeter</strong>: 1 G or less (frequency 500 Hz or<br />
less)<br />
• Remote Flowtube (size 2.5 to 400 mm (0.1 to 16 in.)):<br />
2 G or less (frequency 500 Hz or less)<br />
Note: Avoid locations with much vibration (where the pipe<br />
vibration frequency is 500 Hz or more), which may<br />
cause damage to the equipment.<br />
6-19<br />
IM 01E20D01-01E