8. EXPLOSION PROTECTED TYPE INSTRUMENT<br />

Temperature Code:<br />

Temperature<br />

Code<br />

T6<br />

T5<br />

T4<br />

T3<br />

Maximum Process<br />

Temperature<br />

+70°C (+158°F)<br />

+85°C (+185°F)<br />

+120°C (+248°F)<br />

+150°C (+302°F)<br />

Minimum Process<br />

Temperature<br />

–40°C (–40°F)<br />

–40°C (–40°F)<br />

–40°C (–40°F)<br />

–40°C (–40°F)<br />

T0808.EPS<br />

Ambient Temp.: –40°C to +60°C (–40°F to +140°F)<br />

(2) Installation<br />

For CSA C22.2 Series<br />

WARNING<br />

•All wiring shall comply with Canadian Electrical<br />

Code Part I and Local Electrical Codes.<br />

• In hazardous location, wiring shall be in conduit<br />

as shown in Figure 8.3.1.<br />

WARNING :SEAL ALL CONDUITS WITHIN<br />

50cm OF THE ENCLOSURE’.<br />

UN SCELLEMENT DOIT ÉTRE<br />

INSTALLÉ À MOINS DE 50cm DU<br />

BOÎTIER.<br />

•When installed in Division 2, “SEALS NOT<br />

REQUIRED”<br />

For CSA E79 Series<br />

(<strong>Integral</strong> <strong>Flowmeter</strong>)<br />

WARNING<br />

WARNING :AFTER DE-ENERGIZING, DELAY<br />

20 MINUTES BEFORE OPENING.<br />

APRÉS POWER-OFF, ATTENDRE<br />

20 MINUTES AVANT D’OUVRIR.<br />

• Take care not to generate mechanical spark<br />

when access to the instrument and peripheral<br />

devices in hazardous locations.<br />

(Remote Flowtube)<br />

WARNING<br />

WARNING :DE-ENERGIZE BEFORE OPEN-<br />

ING.<br />

OUVRIR LE CIRCUIT AVANT<br />

D’ENLEVER LE COUVERCLE.<br />

• Take care not to generate mechanical spark<br />

when access to the instrument and peripheral<br />

devices in hazardous locations.<br />

(4) Maintenance and Repair<br />

For CSA E79 Series<br />

WARNING<br />

•All wiring shall comply with local installation<br />

requirements and local electrical code.<br />

• In hazadous locations, the cable entry devices<br />

shall be of a certified flameproof type, suitable<br />

for the conditions of use and correctly installed.<br />

• Unused apertures shall be closed with suitable<br />

flameproof certified blanking elements. (The<br />

plug attached is flameproof certified.)<br />

WARNING<br />

The instrument modification or parts replacement<br />

by other than authorized representative of<br />

YOKOGAWA Electric Corporation or<br />

YOKOGAWA Corporation of AMERICA is<br />

prohibited and will void Canadian Standards<br />

Explosionproof Certification.<br />

HAZARDOUS LOCATIONS<br />

Conduit<br />

<strong>Magnetic</strong> <strong>Flowmeter</strong><br />

Sealing Fitting<br />

(3) Operation<br />

For CSA C22.2 Series<br />

WARNING<br />

WARNING :OPEN CIRCUIT BEFORE REMOV-<br />

ING COVER.<br />

OUVRIR LE CIRCUIT AVANT<br />

D’ENLEVER LE COUVERCLE.<br />

• Take care not to generate mechanical spark<br />

when access to the instrument and peripheral<br />

devices in hazardous locations.<br />

50 cm Max.<br />

Figure 8.3.1 Conduit Wiring<br />

F0803.EPS<br />

8-6<br />

IM 01E20D01-01E