5.5.2 Setting of Write Protect Switch<br />

By setting the write protect function to “Protect” it is<br />

possible to prevent the overwriting of parameters.<br />

Write protection can be carried out using either the<br />

hardware switch on the CPU board (i.e., Switch 2) or<br />

software parameter settings. If either of these items is<br />

set to “Protect,” the overwriting of parameters will be<br />

prohibited.<br />

NOTE<br />

• If the hardware switch is set to “Protect,” it will<br />

not be possible to overwrite parameters;<br />

furthermore, this condition will be maintained<br />

until the switch is set to “Enable.”<br />

• In the case of FOUNDATION Fieldbus protocol,<br />

write protect setting switch is always be<br />

set as “Enable”.<br />

For more details regarding usage of the write protect<br />

function and the software’s parameter switches, refer to<br />

“Chapter 6: Parameter Description” in the user’s<br />

manual of the <strong>AXF</strong> <strong>Integral</strong> <strong>Flowmeter</strong> [Software<br />

Edition] (IM 01E20C02-01E).<br />

5. MAINTENANCE<br />

5.6 Regular Inspection<br />

Items<br />

(1) Inspection of moisture-proofing inside the<br />

terminal box: Once/year<br />

(2) Retightening of piping joint screws: About<br />

twice/year<br />

(3) Inspection of electrodes and lining (in<br />

case of adhesive and/or abrasive fluids,<br />

etc.)<br />

Determine the period of regular inspection as<br />

necessary.<br />

5.7 Excitation Coil and<br />

Insulation Resistance<br />

Check (Remote<br />

Flowtube Only)<br />

WARNING<br />

• Before checking of the excitation coil and the<br />

insulation resistance, be sure that the power<br />

supply for <strong>AXF</strong>A11 or the <strong>AXF</strong>A14 converter<br />

has been turned off.<br />

• Before checking, be sure to disconnect the<br />

cables from the terminals of the remote<br />

flowtube.<br />

(1) Excitation Coil Check (Remote Flowtube<br />

Only)<br />

Check that there is continuity between terminals EX1<br />

and EX2 in the terminal box. If there is no continuity,<br />

the coils may be broken and replacement or repair of<br />

the flowtube is necessary. The coil resistance is<br />

designed to be 150 or less. If it is not, this may be<br />

an abnormal condition. Consult <strong>Yokogawa</strong>’s sales or<br />

service offices.<br />

(2) Insulation Resistance Check (Remote<br />

Flowtube Only)<br />

Check the insulation resistances in the terminal box in<br />

accordance with the tables below. If any of them falls<br />

below the values listed in the tables, consult<br />

<strong>Yokogawa</strong>’s sales or service offices for investigation.<br />

If the insulation resistance cannot be restored, replacement<br />

or repair of the flowtube is needed. In case of<br />

submersible type flowmeters, undo the wiring connection<br />

on the converter side and measure resistance at the<br />

cable terminals.<br />

5-10<br />

IM 01E20D01-01E