5.5.2 Setting of Write Protect Switch<br />
By setting the write protect function to “Protect” it is<br />
possible to prevent the overwriting of parameters.<br />
Write protection can be carried out using either the<br />
hardware switch on the CPU board (i.e., Switch 2) or<br />
software parameter settings. If either of these items is<br />
set to “Protect,” the overwriting of parameters will be<br />
prohibited.<br />
NOTE<br />
• If the hardware switch is set to “Protect,” it will<br />
not be possible to overwrite parameters;<br />
furthermore, this condition will be maintained<br />
until the switch is set to “Enable.”<br />
• In the case of FOUNDATION Fieldbus protocol,<br />
write protect setting switch is always be<br />
set as “Enable”.<br />
For more details regarding usage of the write protect<br />
function and the software’s parameter switches, refer to<br />
“Chapter 6: Parameter Description” in the user’s<br />
manual of the <strong>AXF</strong> <strong>Integral</strong> <strong>Flowmeter</strong> [Software<br />
Edition] (IM 01E20C02-01E).<br />
5. MAINTENANCE<br />
5.6 Regular Inspection<br />
Items<br />
(1) Inspection of moisture-proofing inside the<br />
terminal box: Once/year<br />
(2) Retightening of piping joint screws: About<br />
twice/year<br />
(3) Inspection of electrodes and lining (in<br />
case of adhesive and/or abrasive fluids,<br />
etc.)<br />
Determine the period of regular inspection as<br />
necessary.<br />
5.7 Excitation Coil and<br />
Insulation Resistance<br />
Check (Remote<br />
Flowtube Only)<br />
WARNING<br />
• Before checking of the excitation coil and the<br />
insulation resistance, be sure that the power<br />
supply for <strong>AXF</strong>A11 or the <strong>AXF</strong>A14 converter<br />
has been turned off.<br />
• Before checking, be sure to disconnect the<br />
cables from the terminals of the remote<br />
flowtube.<br />
(1) Excitation Coil Check (Remote Flowtube<br />
Only)<br />
Check that there is continuity between terminals EX1<br />
and EX2 in the terminal box. If there is no continuity,<br />
the coils may be broken and replacement or repair of<br />
the flowtube is necessary. The coil resistance is<br />
designed to be 150 or less. If it is not, this may be<br />
an abnormal condition. Consult <strong>Yokogawa</strong>’s sales or<br />
service offices.<br />
(2) Insulation Resistance Check (Remote<br />
Flowtube Only)<br />
Check the insulation resistances in the terminal box in<br />
accordance with the tables below. If any of them falls<br />
below the values listed in the tables, consult<br />
<strong>Yokogawa</strong>’s sales or service offices for investigation.<br />
If the insulation resistance cannot be restored, replacement<br />
or repair of the flowtube is needed. In case of<br />
submersible type flowmeters, undo the wiring connection<br />
on the converter side and measure resistance at the<br />
cable terminals.<br />
5-10<br />
IM 01E20D01-01E