General Information<br />

Glossary<br />

Refer to Figure I-3.<br />

Inlet Valve – Accepts the sample stream via 1/4 inch NPTF connector.<br />

The user must supply the sample with a pressure between 14 and 100 psig.<br />

Inlet Filter – 60-micron stainless steel filter traps particulate matter in<br />

sample stream.<br />

Bypass/Needle Valve – Used to redirect flow in bypass system.<br />

Pressure Regulator – Adjusts flow on incoming sample stream.<br />

Flow Meter – Measures sample flow rate.<br />

Check Valve – Prevents the backflow of sample.<br />

Restrictor Tube Assembly – Used in conjunction with pressure regulator<br />

to lower downstream pressure.<br />

Reagent Manifold – Directs sample flow in and out of reagent bottle<br />

assembly.<br />

Reagent Bottle Adapter Assembly – Connects reagent bottle assembly to<br />

manifold.<br />

Diffusion Tubing Assembly – Semi-permeable tubing through which<br />

reagent diffuses into the sample.<br />

Reagent Bottle – Contains formic acid reagent that lowers the sample pH<br />

to below 4.<br />

Flow Cell – Contains reference electrode, sensing electrode and ATC probe.<br />

Diverter Valve – Allows the flow cell reservoir to fill during calibration by<br />

forming a closed-loop system.<br />

<strong>Ammonia</strong> Electrode – Senses ammonia ions in sample stream and produce<br />

an electrical potential dependent on sample concentration.<br />

Reference Electrode – Provides a constant reference potential and<br />

completes the measurement circuit.<br />

Reference Electrode Filling Solution Bottle – Provides constant flow of<br />

electrolyte solution through reference electrode for maximum stability.<br />