Instrument Operation<br />

Alarm<br />

mAmp<br />

Set up to 3 alarms: high, low, and an error signaling contact.<br />

• Relays 1 and 2 (RLY1, RLY2) are normally open dry contacts used to set<br />

high and low alarms for measurement values<br />

• RLY 1 options: Off, Hi, Low<br />

• RLY 2 options: Off, Hi, Low<br />

• Relay 3 (RLY3) is a normally closed contact that can be dedicated to<br />

errors. This will close if power to monitor is lost. This alarm is effected<br />

by calibration, errors and offline or hold status<br />

• RLY 3 options: Off, Cal, Hold, Err<br />

• Default: All alarms (1, 2, & 3) are set to off<br />

Set either of 2 analog current outputs (Sout & Tout), set scale values as<br />

linear or logarithmic curves for concentration.<br />

Analog outputs:<br />

• Toggle between 4-20 mA or 0-20 mA current signals.<br />

• The outputs share a common return, but are isolated from the main<br />

circuitry of the monitor.<br />

• Default output current: 4-20 mA<br />

• Sout: dedicated ammonia output.<br />

• Tout: dedicated temperature output.<br />

• Toggle between logarithmic (LOG) and linear (LIn) scale for each.<br />

• Set Low (LO) and high (HI) limits for Sout and Tout.<br />

• Default scale values: for both Log and Lin.<br />

Default Concentration:<br />

• Low limit: 1.0 ppb<br />

• High limit: 100 ppb<br />

Default Temperature:<br />

• Low limit: 5.0 degrees C<br />

• High limit: 45.0 degrees C<br />

mADJ<br />

Set the mAmp offset adjustment value.<br />

• Select the Sout or Tout output.<br />

• Toggle the numeric offset value and positive or negative offset value.<br />

• Default: 00.00 mAmp.<br />

<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />

III-7