2110XP Ammonia Analyzer User Guide (1574 Kb) - Thermo Scientific
2110XP Ammonia Analyzer User Guide (1574 Kb) - Thermo Scientific
2110XP Ammonia Analyzer User Guide (1574 Kb) - Thermo Scientific
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Instrument Operation<br />
Alarm<br />
mAmp<br />
Set up to 3 alarms: high, low, and an error signaling contact.<br />
• Relays 1 and 2 (RLY1, RLY2) are normally open dry contacts used to set<br />
high and low alarms for measurement values<br />
• RLY 1 options: Off, Hi, Low<br />
• RLY 2 options: Off, Hi, Low<br />
• Relay 3 (RLY3) is a normally closed contact that can be dedicated to<br />
errors. This will close if power to monitor is lost. This alarm is effected<br />
by calibration, errors and offline or hold status<br />
• RLY 3 options: Off, Cal, Hold, Err<br />
• Default: All alarms (1, 2, & 3) are set to off<br />
Set either of 2 analog current outputs (Sout & Tout), set scale values as<br />
linear or logarithmic curves for concentration.<br />
Analog outputs:<br />
• Toggle between 4-20 mA or 0-20 mA current signals.<br />
• The outputs share a common return, but are isolated from the main<br />
circuitry of the monitor.<br />
• Default output current: 4-20 mA<br />
• Sout: dedicated ammonia output.<br />
• Tout: dedicated temperature output.<br />
• Toggle between logarithmic (LOG) and linear (LIn) scale for each.<br />
• Set Low (LO) and high (HI) limits for Sout and Tout.<br />
• Default scale values: for both Log and Lin.<br />
Default Concentration:<br />
• Low limit: 1.0 ppb<br />
• High limit: 100 ppb<br />
Default Temperature:<br />
• Low limit: 5.0 degrees C<br />
• High limit: 45.0 degrees C<br />
mADJ<br />
Set the mAmp offset adjustment value.<br />
• Select the Sout or Tout output.<br />
• Toggle the numeric offset value and positive or negative offset value.<br />
• Default: 00.00 mAmp.<br />
<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />
III-7