Chapter IV Calibration<br />
Calibration<br />
Setup<br />
The <strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> ammonia monitor provides simple<br />
and fast calibration cycles using double known addition that quickly have<br />
your system back online with security and confidence. The Orion <strong>2110XP</strong><br />
monthly calibration is quick and accurate using a simple pipette to inject<br />
two known standards into the sample matrix.<br />
Prior to performing a calibration, refer to Chapter III, Use of Setup Mode<br />
for instructions on how to properly configure the Orion <strong>2110XP</strong>. The<br />
value of the calibration standards, volume of the calibration standards, and<br />
the volume of the flow cell must be entered in the setup mode.<br />
Adjust the flow cell volume according to the precise factory calibration<br />
volume that is printed on the side of each flow cell.<br />
<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />