2110XP Ammonia Analyzer User Guide (1574 Kb) - Thermo Scientific
2110XP Ammonia Analyzer User Guide (1574 Kb) - Thermo Scientific
2110XP Ammonia Analyzer User Guide (1574 Kb) - Thermo Scientific
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Instrument Operation<br />
Password protection<br />
Default password: 0000 – indicates password protection has not been<br />
activated.<br />
System password: Management secured password protection of setup and<br />
calibration process.<br />
Calibration password: <strong>User</strong> secured password for protection of calibration<br />
process only.<br />
If password(s) are activated:<br />
• System prompts user to enter system password.<br />
• Marquee: ENTER PASSWORD<br />
• Main display top: PASS<br />
• Main display bottom: 0000 – flashing<br />
• Correct password: allows user to enter setup mode for custom<br />
programming options.<br />
• Incorrect password: Password incorrect or not entered correctly.<br />
• System password<br />
• Marquee: SYSTEM PASS INCORRECT<br />
• Main display: E035.<br />
• Calibration password<br />
• Marquee: CAL PASS INCORRECT<br />
• Main display: E034.<br />
• Verify password and re-enter password.<br />
If password(s) are de-activated:<br />
• System enters the last setup menu screen used by the user.<br />
• Marquee: Flashes current menu selection.<br />
• Main display: SEL SCrn<br />
<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />
III-4