Instrument Operation<br />

Default <strong>User</strong> Action Marquee<br />

Scrolling<br />

Main Display<br />

top line<br />

bottom line<br />

Notes<br />

TEST<br />

TEST<br />

(flashing)<br />

SEL<br />

SCrn<br />

Press to test mAmp outputs (4-<br />

20), relays (rLY), and the air pump (AIr)<br />

DIAGNOSTICS<br />

appears above<br />

SETUP in mode<br />

window<br />

To Select: tOUt<br />

4-20<br />

Use / to scroll through<br />

4-20, rLY, and AIr setting. Select 4-20<br />

SELECT mA OR RELAY<br />

tESt<br />

4-20<br />

(Flashing)<br />

Press<br />

to test 4-20 outputs<br />

Use / to toggle between<br />

SOUt and tOUt. Select tOUt<br />

SELECT TEMP OR SENSOR OUTPUT<br />

SELECT TEMP OR SENSOR OUTPUT<br />

tOUt or SOUt<br />

(Flashing)<br />

tOUt<br />

(Flashing)<br />

Press<br />

for tOUt; low value (LO)<br />

Actual<br />

LO Value<br />

Displayed<br />

mA TEMP OUTPUT LOW VALUE<br />

LO<br />

#. # C<br />

4.0 mA are sourced<br />

at output terminal<br />

11<br />

Press<br />

for tOUt; high value (HI)<br />

Actual<br />

HI Value<br />

Displayed<br />

mA TEMP OUTPUT HIGH VALUE<br />

HI<br />

#. # C<br />

20 mA are sourced<br />

at output terminal<br />

11<br />

Press to return user to main<br />

setup menu for TEST submenu<br />

TEST<br />

(Flashing)<br />

SEL<br />

SCrn<br />

Press to test mAmp outputs<br />

(4-20), relays (rLY), and the air pump<br />

(AIr)<br />

<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />