Troubleshooting<br />
Displayed<br />
Event Code<br />
Description Cause Troubleshooting<br />
System Event/Error Codes - Yellow LED<br />
E004 DEFAULT VALUES Instrument has been reset or is new<br />
out of box<br />
Calibrate the instrument<br />
E021 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />
E101 TEMP OUT OF RANGE Temperature sensor on channel 1 out<br />
of range (5.1 ˚C to 45 ˚C)<br />
E111 CAL DUE <strong>User</strong> calibration required. No<br />
calibration has been performed in<br />
user specified time interval.<br />
E127 SOUT mA OVER RANGE Measured ammonia value above Sout<br />
high limit value<br />
E128 SOUT mA UNDER RANGE Measured ammonia value below Sout<br />
low limit value<br />
E129 TOUT mA OVER RANGE Measured temperature value above<br />
Tout high limit value<br />
E130 TOUT mA UNDER RANGE Measured temperature value below<br />
Tout low limit value<br />
Check ATC cable and probe connections. Replace if needed.<br />
Recalibrate the instrument.<br />
Verify user selectable Sout high limit value in setup menu<br />
Verify user selectable Sout low limit value in setup menu<br />
Verify user selectable Tout high limit value in setup menu<br />
Verify user selectable Tout low limit value in setup menu<br />
System Error Codes - Red LED<br />
E001 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />
E002 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />
E005 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />
E006 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />
E018 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />
E019 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />
E020 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />
E022 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />
E024 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />
E026 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />
E027 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />
E029 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />
E030 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />
E038 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />
<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />
VI-15