Troubleshooting<br />

Displayed<br />

Event Code<br />

Description Cause Troubleshooting<br />

System Event/Error Codes - Yellow LED<br />

E004 DEFAULT VALUES Instrument has been reset or is new<br />

out of box<br />

Calibrate the instrument<br />

E021 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />

E101 TEMP OUT OF RANGE Temperature sensor on channel 1 out<br />

of range (5.1 ˚C to 45 ˚C)<br />

E111 CAL DUE <strong>User</strong> calibration required. No<br />

calibration has been performed in<br />

user specified time interval.<br />

E127 SOUT mA OVER RANGE Measured ammonia value above Sout<br />

high limit value<br />

E128 SOUT mA UNDER RANGE Measured ammonia value below Sout<br />

low limit value<br />

E129 TOUT mA OVER RANGE Measured temperature value above<br />

Tout high limit value<br />

E130 TOUT mA UNDER RANGE Measured temperature value below<br />

Tout low limit value<br />

Check ATC cable and probe connections. Replace if needed.<br />

Recalibrate the instrument.<br />

Verify user selectable Sout high limit value in setup menu<br />

Verify user selectable Sout low limit value in setup menu<br />

Verify user selectable Tout high limit value in setup menu<br />

Verify user selectable Tout low limit value in setup menu<br />

System Error Codes - Red LED<br />

E001 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />

E002 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />

E005 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />

E006 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />

E018 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />

E019 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />

E020 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />

E022 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />

E024 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />

E026 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />

E027 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />

E029 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />

E030 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />

E038 HARDWARE FAULT System Error Contact Technical Support 1-800-225-1480<br />