06.11.2014 Views

2110XP Ammonia Analyzer User Guide (1574 Kb) - Thermo Scientific

2110XP Ammonia Analyzer User Guide (1574 Kb) - Thermo Scientific

2110XP Ammonia Analyzer User Guide (1574 Kb) - Thermo Scientific

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Calibration<br />

Rinsing the Flow Cell<br />

Warning: Be sure to follow the procedure for rinsing the flow cell.<br />

D<br />

C<br />

B<br />

A<br />

50<br />

40<br />

30<br />

20<br />

10<br />

1. Open the sample inlet valve, see A .<br />

• Positioning valve with pointed end facing up for open position.<br />

2. Open the bypass/needle valve by turning counterclockwise, see B .<br />

• Check for appropriate sample flow.<br />

• Adjust pressure regulator if required. See step 3.<br />

3. Adjust the knob on the pressure regulator, see C , so the ball in<br />

the flow meter indicates the correct flow rate reading, see D .<br />

The flow rate should be adjusted to 40 mL/min.<br />

• Pull out the knob to unlock the setting and allow adjustment of the<br />

flow rate.<br />

• Turn the knob clockwise to increase the flow rate.<br />

• Turn the knob counterclockwise to decrease the flow rate.<br />

• Push in the knob to lock the setting.<br />

4. Push in the diverter valve on the flow cell.<br />

• Flow cell will begin to fill.<br />

5. Flush the flow cell for 10 to 60 minutes. The sample will drain from<br />

the siphon tube into the siphon drain to the waste.<br />

6. Begin a DKA calibration. Refer to Chapter IV, Before Performing<br />

DKA Calibration.<br />

Figure IV-2<br />

<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />

IV-3

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!