2110XP Ammonia Analyzer User Guide (1574 Kb) - Thermo Scientific
2110XP Ammonia Analyzer User Guide (1574 Kb) - Thermo Scientific
2110XP Ammonia Analyzer User Guide (1574 Kb) - Thermo Scientific
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Calibration<br />
Rinsing the Flow Cell<br />
Warning: Be sure to follow the procedure for rinsing the flow cell.<br />
D<br />
C<br />
B<br />
A<br />
50<br />
40<br />
30<br />
20<br />
10<br />
1. Open the sample inlet valve, see A .<br />
• Positioning valve with pointed end facing up for open position.<br />
2. Open the bypass/needle valve by turning counterclockwise, see B .<br />
• Check for appropriate sample flow.<br />
• Adjust pressure regulator if required. See step 3.<br />
3. Adjust the knob on the pressure regulator, see C , so the ball in<br />
the flow meter indicates the correct flow rate reading, see D .<br />
The flow rate should be adjusted to 40 mL/min.<br />
• Pull out the knob to unlock the setting and allow adjustment of the<br />
flow rate.<br />
• Turn the knob clockwise to increase the flow rate.<br />
• Turn the knob counterclockwise to decrease the flow rate.<br />
• Push in the knob to lock the setting.<br />
4. Push in the diverter valve on the flow cell.<br />
• Flow cell will begin to fill.<br />
5. Flush the flow cell for 10 to 60 minutes. The sample will drain from<br />
the siphon tube into the siphon drain to the waste.<br />
6. Begin a DKA calibration. Refer to Chapter IV, Before Performing<br />
DKA Calibration.<br />
Figure IV-2<br />
<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />
IV-3