Calibration<br />
Before Performing<br />
DKA Calibration<br />
Recommendations<br />
Maximum system accuracy is ensured through a fast, easy, and accurate<br />
calibration performed in the expected sample range. This calibration<br />
procedure uses equipment supplied in the calibration kit (Cat. No.<br />
<strong>2110XP</strong>CAL).<br />
• Wait at least one hour after changing reagent.<br />
• Check that the present sample concentration is equal or less than<br />
standard 1; if not, calibration can still be performed with a loss in<br />
accuracy as long as background concentration is not significantly larger.<br />
The best results are achieved when standard 1 is twice the level of the<br />
present sample concentration.<br />
• Proper pipette technique is crucial to calibration. If user is not familiar<br />
with pipette equipment or technique, refer to Appendix A-11, Pipette<br />
Operation.<br />
• Do not touch pipette tip with bare fingers. Risk of contamination is<br />
possible. Use powder free gloves to prevent contamination.<br />
• When using a pipette to inject standards during calibration, use a fresh<br />
pipette tip for each standard and each injection.<br />
• When a fresh tip is installed, rinse the tip with the same concentration<br />
standard as the intended injection.<br />
• Fill the tip and dispense the standard 1-3 times before injecting the<br />
standard into the flow cell. Be sure to discard the standard in the<br />
drain between the rinses.<br />
• Wipe any excess liquid from the outside of the tip with out touching<br />
the opening.<br />
• Insert the pipette tip into calibration port on flow cell cap.<br />
Figure IV-4<br />
Pipette Dispensing<br />
• Be sure to hold the pipette completely vertical while dispensing. Do<br />
not inject the standard on the side of the flow cell, as a film may be<br />
retained on the inside of the wall, resulting in errors in precision of<br />
calibration. Refer to Figure IV-4.<br />