Calibration<br />

Before Performing<br />

DKA Calibration<br />

Recommendations<br />

Maximum system accuracy is ensured through a fast, easy, and accurate<br />

calibration performed in the expected sample range. This calibration<br />

procedure uses equipment supplied in the calibration kit (Cat. No.<br />

<strong>2110XP</strong>CAL).<br />

• Wait at least one hour after changing reagent.<br />

• Check that the present sample concentration is equal or less than<br />

standard 1; if not, calibration can still be performed with a loss in<br />

accuracy as long as background concentration is not significantly larger.<br />

The best results are achieved when standard 1 is twice the level of the<br />

present sample concentration.<br />

• Proper pipette technique is crucial to calibration. If user is not familiar<br />

with pipette equipment or technique, refer to Appendix A-11, Pipette<br />

Operation.<br />

• Do not touch pipette tip with bare fingers. Risk of contamination is<br />

possible. Use powder free gloves to prevent contamination.<br />

• When using a pipette to inject standards during calibration, use a fresh<br />

pipette tip for each standard and each injection.<br />

• When a fresh tip is installed, rinse the tip with the same concentration<br />

standard as the intended injection.<br />

• Fill the tip and dispense the standard 1-3 times before injecting the<br />

standard into the flow cell. Be sure to discard the standard in the<br />

drain between the rinses.<br />

• Wipe any excess liquid from the outside of the tip with out touching<br />

the opening.<br />

• Insert the pipette tip into calibration port on flow cell cap.<br />

Figure IV-4<br />

Pipette Dispensing<br />

• Be sure to hold the pipette completely vertical while dispensing. Do<br />

not inject the standard on the side of the flow cell, as a film may be<br />

retained on the inside of the wall, resulting in errors in precision of<br />

calibration. Refer to Figure IV-4.<br />

<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />

IV-5