Instrument Maintenance<br />

Replacement of Restrictor<br />

Tubing<br />

Use the restrictor tube assembly (Cat. No. 2100RT). Refer to Figure V-3<br />

for restrictor tubing connections.<br />

1. Close the inlet valve to stop sample flow through monitor.<br />

2. Loosen the restrictor tube fitting connection from the top of the flow<br />

meter.<br />

3. Loosen the restrictor tube fitting from the left side of the flow cell<br />

block.<br />

4. Pull out the restrictor tube assembly at each end.<br />

5. Unscrew the clip behind flow cell that holds the restrictor tube to panel.<br />

6. Replace with new restrictor tubing assembly.<br />

7. Be sure all fittings are secure, to prevent leaks.<br />

8. Open the inlet valve to start sample flow through monitor.<br />

restrictor<br />

tube assembly<br />

restrictor<br />

adapter<br />

fitting<br />

clamp<br />

flow<br />

meter<br />

O-ring<br />

Figure V-3<br />

Restrictor Tubing<br />

to check valve on<br />

flow cell manifold<br />

<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />

V-6