Calibration<br />
Stage <strong>User</strong> Action Marquee Main Display<br />
top line<br />
bottom line<br />
Notes<br />
8 When the liquid level in the flow cell<br />
begins to drop, turn the inlet valve<br />
off.<br />
Open the calibration port on the flow<br />
cell cap.<br />
WHEN LIQUID LEVEL BEGINS<br />
TO DROP TURN INLET VALVE OFF<br />
OPEN CAL PORT<br />
PRESS ENTER<br />
FLO OFF<br />
9 Press .<br />
10 Wait 2 minutes. STARTING CALIBRATION<br />
MEASURING BACKGROUND<br />
WAIT 2 MINUTES<br />
Actual readings<br />
flashing<br />
System circulates sample.<br />
Minutes will be counted<br />
down on Marquee.<br />
11 Wait 0 - 8 minutes. AWAITING DATA STABILITY Actual readings<br />
flashing<br />
System continuously<br />
evaluates data by applying<br />
stability criteria until a<br />
“stable/pass” condition is<br />
met. If system stabilizes<br />
follow stage 12.<br />
If system stability is not<br />
reached a timeout occurs.<br />
Error E041 is issued. Jump<br />
to step Error E041.<br />
12 Add standard 1 through the<br />
calibration port.<br />
ADD STD1<br />
PRESS ENTER<br />
Add Std1<br />
Pipette 0.5 mL of standard<br />
1 into the flow cell using<br />
the 0.5 mL pipette<br />
13 Press .<br />
14 Wait 2 minutes. MEASURING STD1<br />
CIRCULATING<br />
WAIT 2 MINUTES<br />
Actual readings<br />
flashing<br />
System circulates sample.<br />
Minutes will be counted<br />
down on Marquee.<br />
15 Wait 0 - 8 minutes. AWAITING DATA STABILITY Actual readings<br />
flashing<br />
System continuously<br />
evaluates data by applying<br />
stability criteria until a<br />
“stable/pass” condition is<br />
met. If system stabilizes<br />
follow stage 16.<br />
If system stability is not<br />
reached a timeout occurs.<br />
Error E041 is issued. Jump<br />
to step Error E041.<br />
<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />
IV-7