Troubleshooting<br />
Resetting the<br />
Monitor<br />
The <strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> ammonia monitor allows the user to<br />
reset the monitor through the setup mode or by a hard reset. Resetting the<br />
monitor will restore all settings in the setup menu to factory default values.<br />
Refer to Appendix A-2, ISE Default Values.<br />
Warning: Resetting the monitor will erase all data logs including<br />
calibration, and password protection settings for setup and calibration.<br />
The instrument will reset setup parameters to factory default settings.<br />
<strong>User</strong> Action<br />
Marquee<br />
Scrolling<br />
Main Display<br />
top line<br />
bottom line<br />
Notes<br />
Press<br />
to enter setup mode.<br />
Last menu used<br />
in setup mode<br />
SEL<br />
SCrn<br />
SETUP appears<br />
as mode<br />
Press / until RSET<br />
appears in the Marquee.<br />
RSET<br />
(Flashing)<br />
SEL<br />
SCrn<br />
Press .<br />
Press in the following sequence:<br />
1.<br />
2.<br />
3.<br />
System resets to default values<br />
and returns to the measurement<br />
mode. Actual values for<br />
concentration and temperature<br />
will be displayed if the electrodes<br />
are operational.<br />
PRESS TEST<br />
VIEW ENTER TO<br />
RESET<br />
PRESS TEST<br />
VIEW ENTER TO<br />
RESET<br />
rSEt<br />
?<br />
rSEt<br />
?<br />
<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />
VI-17