Calibration<br />
Span Check<br />
Procedure<br />
The <strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> ammonia monitor remains in<br />
sample mode during the entire procedure.<br />
1. Fill flow cell to calibration level (V 0 , approximately 95 mL).<br />
2. Note concentration displayed when reading is reasonably stable (C 0 ).<br />
3. Add standard 1, volume V 1 , with concentration C 1 .<br />
4. When reading is reasonably stable, note concentration (C S ).<br />
5. The noted value should be within ± 10% of the calculated value<br />
according to:<br />
C S = (C 0 x V 0 + C 1 x V 1 ) / (V 0 + V 1 )<br />
• C 1 should be ammonia concentration actually in the standard<br />
bottle.<br />
<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />
IV-13