2110XP Ammonia Analyzer User Guide (1574 Kb) - Thermo Scientific
2110XP Ammonia Analyzer User Guide (1574 Kb) - Thermo Scientific
2110XP Ammonia Analyzer User Guide (1574 Kb) - Thermo Scientific
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Display Test<br />
• Press to start display test.<br />
1. All the display segments will light up and the system will cycle<br />
through the LED colors.<br />
2. Verify that all the display segments are on and that the LED colors<br />
are functional and press .<br />
3. All the display segments will turn off except for the Marquee display,<br />
which should show PRESS ENTER TO CONTINUE.<br />
4. Verify that all the display segments are off and press .<br />
5. The system will show a counting number pattern on both lines of<br />
the main display.<br />
6. Verify that the display segments are functional and press to<br />
proceed to the keypad test.<br />
Menu <strong>User</strong> Action Marquee Scrolling Main Display<br />
top line<br />
bottom line<br />
Notes<br />
Display Test<br />
PRESS ENTER FOR<br />
DISPLAY TEST<br />
dISP<br />
tESt<br />
Press .<br />
Verify that all the display segments<br />
are on and that the LED colors are<br />
functional.<br />
All segments on All segments on System will cycle through<br />
the LED colors<br />
Press .<br />
Verify that all the display segments<br />
are off.<br />
PRESS ENTER TO<br />
CONTINUE<br />
All segments off<br />
Press .<br />
Verify that the display segments are<br />
functional.<br />
PRESS ENTER TO<br />
CONTINUE<br />
Count pattern<br />
Count pattern<br />
Press .<br />
<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />
VI-7