Instrument Preparation<br />

Terminal<br />

Assignments<br />

Figure II-6<br />

Terminal Assignments<br />

Alarms / Relays Analog Outputs Temperature ISE<br />

1 Relay 1<br />

(normally open)<br />

2 Relay 1<br />

(normally open)<br />

3 Relay 2<br />

(normally open)<br />

4 Relay 2<br />

(normally open)<br />

5 Relay 3<br />

(normally closed)<br />

6 Relay 3<br />

(normally closed)<br />

7 Air pump<br />

9 Sout (mA)<br />

<strong>Ammonia</strong> Signal<br />

18 ATC 21-<br />

22<br />

Not Used<br />

10 GND<br />

19 ATC 23 Electrode<br />

Common Ground * Shield **<br />

11 Tout (mA)<br />

Temp. Signal<br />

12-<br />

17<br />

20 Solution<br />

Ground<br />

24 Reference<br />

Electrode<br />

Not Used 25 Not Used<br />

26 <strong>Ammonia</strong><br />

Electrode<br />

8 Air pump<br />

Key Electrical Notes for the Orion <strong>2110XP</strong>:<br />

• Fluid ground is provided through the Orion ATC Probe (30 kΩ),<br />

Cat. No. 2100TP. Only an Orion ATC probe can be used to provide<br />

the fluid ground.<br />

• Do not connect to any of the unused terminals: 12-17, 21-22 and 25.<br />

* Terminal 9 & 11 are not isolated from each other. They share terminal 10 as a<br />

common ground.<br />

** Do not attach external ground shield connections to terminal 23<br />

<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />