Instrument Preparation<br />
Sample Requirements<br />
Additional information is listed in Appendix A-3, Specifications.<br />
Sample inlet connection – 1/4” NPTF. If particulate matter is present<br />
in sample, pre-filtration is necessary. The 60-micron stainless steel filter<br />
located after inlet valve will remove moderate amounts of particulates.<br />
Flow rate – 40 mL/min (nominal).<br />
Pressure – 8-100 psig. Consult Technical Support* for details on sample<br />
handling if pressure is outside range.<br />
Temperature – Temperature must be between 5-35 ˚C.<br />
<strong>Ammonia</strong> level – <strong>Ammonia</strong> levels are read directly in ppb or ppm, when<br />
calibrated with Orion standards 1 and 2.<br />
Sample alkalinity – Sample alkalinity should not be more than 250<br />
ppm CaCO 3 equivalent. For higher sample alkalinity, contact Technical<br />
Support*.<br />
Sulfite – Sulfite cannot be present in the sample.<br />
* Technical Support<br />
US & Canada: 800-225-1480<br />
Dom. Fax: 978-232-6015<br />
Int’l. Fax: 978-232-6031<br />
www.thermo.com/water<br />
<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />