Instrument Preparation<br />

Sample Requirements<br />

Additional information is listed in Appendix A-3, Specifications.<br />

Sample inlet connection – 1/4” NPTF. If particulate matter is present<br />

in sample, pre-filtration is necessary. The 60-micron stainless steel filter<br />

located after inlet valve will remove moderate amounts of particulates.<br />

Flow rate – 40 mL/min (nominal).<br />

Pressure – 8-100 psig. Consult Technical Support* for details on sample<br />

handling if pressure is outside range.<br />

Temperature – Temperature must be between 5-35 ˚C.<br />

<strong>Ammonia</strong> level – <strong>Ammonia</strong> levels are read directly in ppb or ppm, when<br />

calibrated with Orion standards 1 and 2.<br />

Sample alkalinity – Sample alkalinity should not be more than 250<br />

ppm CaCO 3 equivalent. For higher sample alkalinity, contact Technical<br />

Support*.<br />

Sulfite – Sulfite cannot be present in the sample.<br />

* Technical Support<br />

US & Canada: 800-225-1480<br />

Dom. Fax: 978-232-6015<br />

Int’l. Fax: 978-232-6031<br />

www.thermo.com/water<br />

<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />

II-3

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!