Calibration<br />
Performing DKA<br />
Calibration<br />
• To begin a DKA calibration, the system starts from the measurement<br />
mode.<br />
• The calibration port must be closed for proper flow cell pressure.<br />
• The value of the calibration standards, volume of the calibration<br />
standards, and the volume of the flow cell must be entered in the setup<br />
menu prior to performing a calibration. Refer to Chapter III, Use of<br />
Setup Mode and Chapter IV, Calibration Setup.<br />
Stage <strong>User</strong> Action Marquee Main Display<br />
top line<br />
bottom line<br />
Notes<br />
1 Actual temperature reading Actual readings in<br />
concentration and mV<br />
2* Press to start calibration.<br />
CAL appears as mode.<br />
* If calibration password is activated<br />
- user must scroll in digits 0-9 using<br />
or to set digits and<br />
to move selected digit.<br />
Press to accept<br />
password.<br />
ENTER PASSWORD<br />
PASS<br />
0000<br />
3 DKA or OFFL<br />
(Flashing)<br />
CAL<br />
tYPE<br />
4<br />
Press / to select DKA.<br />
DKA<br />
(Flashing)<br />
CAL<br />
tYPE<br />
5 Press .<br />
6 Check that the calibration port<br />
on the flow cell cap is closed.<br />
Push the diverter valve in.<br />
7 Press .<br />
CHECK CAL PORT CLOSED<br />
PUSH DIVERTER VALVE IN<br />
PRESS ENTER<br />
FILL<br />
FCEL<br />
<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />
IV-6