Calibration<br />
Stage<br />
<strong>User</strong> Action<br />
Marquee<br />
Main Display<br />
top line<br />
bottom line<br />
Notes<br />
16 Add standard 2 through the<br />
calibration port.<br />
ADD STD2<br />
PRESS ENTER<br />
Add Std2<br />
Pipette 0.5 mL of standard 2 into the<br />
flow cell using the 0.5 mL pipette<br />
17 Press .<br />
18 Wait 2 minutes. MEASURING STD2<br />
CIRCULATING<br />
WAIT 2 MINUTES<br />
Actual readings<br />
flashing<br />
System circulates sample. Minutes<br />
will be counted down on Marquee.<br />
19 Wait 0 - 8 minutes. AWAITING DATA STABILITY Actual readings<br />
flashing<br />
System continuously evaluates data<br />
by applying stability criteria until a<br />
“stable/pass” condition is met. If<br />
system stabilizes follow stage 20.<br />
If system stability is not reached a<br />
timeout occurs. Error E041 is issued.<br />
Jump to step Error E041.<br />
20 No action necessary. System calculates new Eo and slope<br />
and checks for errors. If no errors<br />
are found follow<br />
stage 21.<br />
If there is a slope/E 0 or other<br />
cal data related error, jump to<br />
appropriate error code.<br />
21 Press to accept the<br />
new slope and Eo values<br />
or<br />
Press to abort the<br />
calibration. The new slope<br />
and Eo values will not be<br />
stored. The values for the<br />
last saved calibration prior<br />
to the aborted calibration<br />
will be used in the<br />
measurement mode.<br />
NEW CALIBRATION SLOPE AND E 0<br />
PRESS ENTER TO ACCEPT OR PRESS<br />
EXIT TO ABORT WITHOUT SAVING<br />
Calculated Slope<br />
and E 0<br />
System presents the new slope and<br />
E 0 values.<br />
If is pressed, continue to<br />
follow stage 22.<br />
If is pressed, refer to<br />
Exit 4.<br />
<strong>Thermo</strong> <strong>Scientific</strong> Orion <strong>2110XP</strong> <strong>Ammonia</strong> Monitor <strong>User</strong> <strong>Guide</strong><br />
IV-8