Measurement Value Output7.2 Digital Value Output7.2.1 Data Protocol ILD1401The digital measurement values are issued as unsigned digital values (raw values).Digital value Used for0 ... 39 SMR back-up40 ... 4055 Measurement range4056 ... 4095 EMR back-upCalculation of a measurement value in mm from digital output:Reference value Start of Measuring Range:Reference value Midrangex [mm] = (digital OUT *1.024096- 0.01) * MR [mm]x [mm] = (digital OUT *1.024096- 0.51) * MR [mm]Example: MR =10 mm, digital value = 2048, measurement value = 5 mm or 0 mmNote: A digital value can be calculated from a measurement value (millimeter) as follows:x [mm]digitalOUT= + 0.01 *MR [mm]40961.02<strong>optoNCDT</strong> <strong>1402</strong>Page 38