Default Setting10. Default Setting--Data protocol ILD<strong>1402</strong>, Binary format--Current output with error value (3.75 mA)--Measuring rate : 1.5 kHz--Interface: 115.2 kBaud, binary format (no ASCII)--Moving average avg =1 (no averaging)--Teach value 1: 0.0--Teach value 2: 16368.0--External input for scaling 1--Continuous measurement output--Output 1. measuring value after switching on: 500ms--Key lock after 5 min power on 1--Settings saved into FLASH--Measuring range:• 100 % FSO: I = 20 mA , digital 16207• 0 % FSO: I = 4 mA, digital 161--Maximum output (101 % FSO):20.16 mA / digital 16367--Minimum output (-1 % FSO):3.84 mA / digital 0Set sensor on default settings: 2Switch off the sensors power supply.Keep the key “Select“ pressed.Switch on the sensors power supply.LED on sensor flashes green.Press the key “Select“ again.LED flashes green three times, approx. 1 Hz.During this time the factory parameters are set.Then the sensor re-boots.1) Only with sensor ILD <strong>1402</strong>-x2) Restoring the factory setting with ILD <strong>1402</strong>-xSC is only possible via the serial interface.<strong>optoNCDT</strong> <strong>1402</strong>Page 84