Measurement Value Output7.2.2 Data Protocol ILD<strong>1402</strong>The digital measurement values are issued as unsigned digital values (raw values).Digital value Used for Digital value Used for0 ... 16367 Value range 16208 ... 16367 EMR back-up (1 %)0 ... 160 SMR back-up (1 %) 16370 ... 16383 Error codes161 ... 16207 Measurement rangeCalculation of a measurement value in mm from digital output:Reference value Start of Measuring Range:Reference value Midrange1.02x [mm] = (digital OUT *16368- 0.01) * MR [mm]x [mm] = (digital OUT *1.02- 0.51) * MR [mm]16368Example: MR =10 mm, digital value = 8184, measurement value = 5 mm or 0 mmNote: A digital value can be calculated from a measurement value (millimeter) as follows:x [mm]digitalOUT= + 0.01 *MR [mm]163681.02<strong>optoNCDT</strong> <strong>1402</strong>7.3 Digital Error CodesDigital error codes are issued in the same way as measurement values.Value range for error codes: 16370 ... 16384 (digital ) OUT--16370 no object detected--16372 too close to the sensor--16374 too far from the sensor--16376 target can not be evaluated--16378 external laser off--16380 target moves towards the sensor--16382 target moves away from sensor--16383 internal errorPage 39