11.07.2015 Views

Manual optoNCDT 1402 - Micro-Epsilon

Manual optoNCDT 1402 - Micro-Epsilon

Manual optoNCDT 1402 - Micro-Epsilon

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Serial Interface RS4228.4.17 Peak Detection with Video Signal, SET_PEAKSEARCHINGName:Description:SET_PEAKSEARCHINGSpecifies the search algorithm.Parameter:--X = 0 --> peak with global maximum--X = 1 --> first peak, direction pixel 0 up to pixel 127, left to right--X = 2 --> last peak, direction pixel 0 up to pixel 127, left to rightFormat: 31 24 23 16 15 8 7 0 hex„+“ „+“ „+“ 0x0d („CR“) 0x2B2B2B0D„l“ „L“ „D“ „1“ 0x494C44310x20 0xFB 0x00 0x03 0x20FB00030x00 0x00 0x00 0x0X 0x0000000XReply: 31 24 23 16 15 8 7 0 hex„l“ „L“ „D“ „1“ 0x494C44310xA0 0xFB 0x00 0x02 0xA0FB00020x20 0x20 0x0D 0x0A 0x20200D0AA peak inside the video signal is limited by overrun and subsequent shortfall the threshold limit. Several validpeaks are evaluated when measuring glass. Measurements on metal surfaces can also produce multiplepeaks. Determine the valid peaks in the software tool (video signal).<strong>optoNCDT</strong> <strong>1402</strong>Page 68

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!