CONELOG ® IMPLANT SYSTEM<br />

PROSTHETIC RESTORATIONS<br />

CONELOG ®<br />

BAR ABUTMENT<br />

CONELOG ® IMPRESSION TAKING OPTIONS<br />

After successful implant insertion, the impression can be taken in two<br />

versions:<br />

IMPRESSION TAKING OVER CONELOG ® BAR ABUTMENTS:<br />

Impression taking after final insertion of the CONELOG ® bar abutments<br />

with impression post for bar abutment. The cast is then fabricated with<br />

bar-lab analogs/soldering aids.<br />

IMPRESSION TAKING IN THE CONELOG ® IMPLANT:<br />

Impression taken in the CONELOG ® implant with CONELOG ® impression<br />

post, open or closed tray, before insertion of CONELOG ® bar abutments.<br />

The cast is then fabricated with CONELOG ® lab analogs. See also pages<br />

18–28.<br />

106<br />

IMPRESSION TAKING OVER CONELOG ® BAR<br />

ABUTMENTS<br />

INSERTION OF CONELOG ® BAR ABUTMENTS<br />

After successful implant insertion and determination of the appropriate gingival<br />

height, the CONELOG ® bar abutments are inserted into the CONELOG ®<br />

implants.<br />

The abutments are inserted into the driver for CONELOG ® bar abutments.<br />

A screw integrated in the instrument secures the abutment. The screw is<br />

tightened by hand.<br />

Driver for CONELOG ®<br />

Bar abutments<br />

Ø 3.3/3.8/4.3/5.0 mm<br />

Bar abutment/instrument<br />

connection<br />

Driver for CONELOG ® ART. NO. ARTICLE MATERIAL<br />

C5300.0020 Bar abutments,<br />

for Ø 3.3/3.8/4.3/5.0 mm<br />

Stainless steel<br />

TIGHTENING TORQUE FOR CONELOG ® The CONELOG<br />

BAR ABUTMENTS<br />

Ø 3.3 mm 20 Ncm<br />

Ø 3.8/4.3/5.0 mm 30 Ncm<br />

® bar abutments are inserted into the previously cleaned<br />

CONELOG ® implants and the torque wrench is then used to tighten the<br />

abutments definitely in the implants based on the specified tightening<br />

torque.<br />

CONELOG ® abutments must be retightened with the same torque after<br />

about five minutes to reach the maximum screw tension. This prevents<br />

screws from loosening to the extent possible.