CONELOG ® IMPLANT SYSTEM<br />
PROSTHETIC RESTORATIONS<br />
CONELOG ®<br />
BAR ABUTMENT<br />
CONELOG ® IMPRESSION TAKING OPTIONS<br />
After successful implant insertion, the impression can be taken in two<br />
versions:<br />
IMPRESSION TAKING OVER CONELOG ® BAR ABUTMENTS:<br />
Impression taking after final insertion of the CONELOG ® bar abutments<br />
with impression post for bar abutment. The cast is then fabricated with<br />
bar-lab analogs/soldering aids.<br />
IMPRESSION TAKING IN THE CONELOG ® IMPLANT:<br />
Impression taken in the CONELOG ® implant with CONELOG ® impression<br />
post, open or closed tray, before insertion of CONELOG ® bar abutments.<br />
The cast is then fabricated with CONELOG ® lab analogs. See also pages<br />
18–28.<br />
106<br />
IMPRESSION TAKING OVER CONELOG ® BAR<br />
ABUTMENTS<br />
INSERTION OF CONELOG ® BAR ABUTMENTS<br />
After successful implant insertion and determination of the appropriate gingival<br />
height, the CONELOG ® bar abutments are inserted into the CONELOG ®<br />
implants.<br />
The abutments are inserted into the driver for CONELOG ® bar abutments.<br />
A screw integrated in the instrument secures the abutment. The screw is<br />
tightened by hand.<br />
Driver for CONELOG ®<br />
Bar abutments<br />
Ø 3.3/3.8/4.3/5.0 mm<br />
Bar abutment/instrument<br />
connection<br />
Driver for CONELOG ® ART. NO. ARTICLE MATERIAL<br />
C5300.0020 Bar abutments,<br />
for Ø 3.3/3.8/4.3/5.0 mm<br />
Stainless steel<br />
TIGHTENING TORQUE FOR CONELOG ® The CONELOG<br />
BAR ABUTMENTS<br />
Ø 3.3 mm 20 Ncm<br />
Ø 3.8/4.3/5.0 mm 30 Ncm<br />
® bar abutments are inserted into the previously cleaned<br />
CONELOG ® implants and the torque wrench is then used to tighten the<br />
abutments definitely in the implants based on the specified tightening<br />
torque.<br />
CONELOG ® abutments must be retightened with the same torque after<br />
about five minutes to reach the maximum screw tension. This prevents<br />
screws from loosening to the extent possible.