CONELOG ® IMPLANT SYSTEM<br />

PROSTHETIC RESTORATIONS<br />

CONELOG ®<br />

BAR ABUTMENT<br />

INSERTING THE BAR CONSTRUCTION<br />

The CONELOG ® bar abutments are transferred from the working cast to<br />

the CONELOG ® implants and screwed in with the prescribed torques. The<br />

finished bar construction is transferred to the CONELOG ® bar abutments<br />

and fixed with new unused prosthetic screws, using a screwdriver (hex)<br />

with 15 Ncm. The newly created full denture is then inserted and checked<br />

for proper fit.<br />

SOLDERED BAR CONSTRUCTION<br />

BASE FOR BAR ABUTMENT, SOLDERABLE<br />

Soldered bar construction with prefabricated bar bases made of solderable<br />

gold alloy. A prosthetic screw for bar abutment (hex) is used to fix the bar<br />

base on the CONELOG ® bar abutment. The height of the bar base is 5.3 mm.<br />

EXAMPLE:<br />

WORKING CAST WITH BAR LAB ANALOGS/SOLDERING AIDS<br />

FOR BAR ABUTMENT (STAINLESS STEEL).<br />

The bar elements are cut accordingly and in consideration of a joining gap<br />

that is as small as possible fitted between the bar bases. The bar components<br />

are mounted to residue-free burn-out plastic, the prosthetic screws<br />

loosen after curing and the bar lifted off the cast. Bar lab analogs/soldering<br />

aids (stainless steel) are inserted in the bar bases and hand tightened with<br />

screws, hex, for bar abutments (thread M 1.6).<br />

With the bar prepared for soldering, a soldering model is fabricated in the<br />

conventional manner.<br />

118<br />

Setting up the solderable<br />

bar bases<br />

Fixed bar components<br />

Bar bases with fitted prefabricated<br />

bar components made of solderable<br />

gold alloy<br />

Base for CONELOG ® ART. NO. ARTICLE MATERIAL<br />

C2258.4300<br />

bar abutment, solderable,<br />

Solderable<br />

for Ø 3.3/3.8/4.3/5.0 mm<br />

gold alloy<br />

Bar lab analogs/soldering aids with bar<br />

and screws, hex, for bar abutment<br />

NOTE<br />

The instruction manuals of the soldering material manufactures must be<br />

observed!<br />

ART. NO. ARTICLE MATERIAL<br />

J4005.1610<br />

Screw, hex, for bar abutment,<br />

for impression taking open tray and for soldering aid, thread M 1.6,<br />

for Ø 3.3/3.8/4.3/5.0 mm<br />

Stainless steel