CONELOG ® IMPLANT SYSTEM<br />

PROSTHETIC RESTORATIONS<br />

CONELOG ®<br />

BALL ABUTMENT<br />

FABRICATION OF A NEW BALL ABUTMENT<br />

RETAINED FULL DENTURE WITH INTEGRATED<br />

METAL REINFORCEMENT<br />

SELECTION AND INSERTION OF THE CONELOG ®<br />

BALL ABUTMENT<br />

After fabricating the cast with CONELOG ® lab analogs, the CONELOG ® ball<br />

abutments are inserted. In conformance with the specified implant diameters<br />

and gingival heights, the dental technician selects the CONELOG ® ball<br />

abutments on the master cast and uses the driver for ball abutments to<br />

manually screw them into the CONELOG ® lab analog.<br />

Different abutment heights can be selected to compensate for differences<br />

in levels in the gingival margin and implants. The ball heads should be at a<br />

uniform level for the best possible retention effect. The ball abutment plateau<br />

should be approx. 1 mm supragingival.<br />

ALIGNING THE DUPLICATION AIDS/SPACER<br />

The included stabilizing rings (white) are placed over the CONELOG ® ball<br />

abutments and the red duplication aids/spacer are clipped on. Thereby the<br />

duplication aids/spacer are aligned parallel to the implant axis. If the implant<br />

axes diverge, the stabilizing rings are NOT used. The implant axial<br />

Divergence must not exceed 10° per implant. For a uniform “engaging” of<br />

the matrices and uniform activating of the lamella retention inserts, the<br />

removal and insertion direction of the denture is important. The duplication<br />

aids/spacer are placed parallel on the CONELOG ® ball abutments in the<br />

common insertion direction and stabilized with wax. Alternatively, the matrix<br />

may be used instead of the duplication aid.<br />

126<br />

Possible fabrication of the duplication cast with duplication aids<br />

or matrices CM Dalbo ® Plus<br />

< max 10°<br />