For impression taking, the impression post for the CONELOG ® bar abutment<br />

is inserted into the driver for impression posts and healing caps for<br />

CONELOG ® bar abutments.<br />

The impression post is then screwed onto the CONELOG ® bar abutment<br />

fitted in the CONELOG ® implant.<br />

A closed tray is suitable for impression taking. The impression is then taken<br />

with silicone or polyether impression material.<br />

Holding screw<br />

Driver<br />

Impression post<br />

Driver for impression posts<br />

and healing caps for CONELOG ®<br />

Bar abutments<br />

Mounted impression<br />

post<br />

After removing the impression, the impression posts<br />

remain on the CONELOG ® bar abutments.<br />

CONELOG ® IMPLANT SYSTEM<br />

PROSTHETIC RESTORATIONS<br />

Driver for impression posts and healing caps for CONELOG ®<br />

Bar abutments,<br />

for Ø 3.3/3.8/4.3/5.0 mm<br />

Impression post for CONELOG ® ART. NO. ARTICLE MATERIAL<br />

C5300.0027<br />

Stainless steel<br />

C2130.4300<br />

bar abutment,<br />

for Ø 3.3/3.8/4.3/5.0 mm<br />

Titanium alloy<br />

107