MATERIALS<br />

CONELOG ® IMPLANT SYSTEM<br />

PROSTHETIC RESTORATIONS<br />

TITANIUM GRADE 4 PROPERTIES:<br />

Chemical structure (in %): O 0.4 max.<br />

Fe 0.3 max.<br />

C 0.1 max.<br />

N 0.05 max.<br />

H 0.0125 max.<br />

Ti > 99.0<br />

Mechanical properties: Tensile strength 680 MPa min.<br />

Elongation 10 %<br />

TITANIUM ALLOY TI6AI4V ELI PROPERTIES:<br />

Chemical structure (in %): AI 5.5–6.75 max.<br />

V 3.5–4.5 max.<br />

Fe 0.3 max.<br />

C 0.08 max.<br />

N 0.05 max.<br />

H 0.015 max.<br />

Ti ~ 90<br />

Mechanical properties: Tensile strength 860 MPa min.<br />

Elongation 10 %<br />

CAST-ON GOLD ALLOY PROPERTIES:<br />

CONELOG ® GOLD-PLASTIC<br />

ABUTMENT<br />

Chemical structure (in %): Au 60<br />

Pd 20<br />

Pt 19<br />

Ir 1<br />

Physical properties: Melting range 1400–1490 °C<br />

Density 17.5 g/cm3<br />

E-Modul 136 GPa<br />

Coefficient of thermal expansion (20–500°C) 11.9 μm/m·°C<br />

Coefficient of thermal expansion (20–600°C) 12.2 μm/m·°C<br />

Color white<br />

Mechanical properties: drawn<br />

Hardness HV5 > 215<br />

Tensile strength (Rm) > 750 MPa<br />

0.2% Elongation limit (Rp 0.2%) > 650 MPa<br />

Elongation at break > 2 %<br />