CONELOG ® IMPLANT SYSTEM<br />
PROSTHETIC RESTORATIONS<br />
CONELOG ®<br />
VARIO SR ABUTMENTS<br />
APPLICATION<br />
SELECTION AND INSERTION OF THE CONELOG ®<br />
VARIO SR ABUTMENTS<br />
The suitable CONELOG ® Vario SR abutment can be selected depending on<br />
the clinical situation for bridge and bar constructions on the patient directly.<br />
CONELOG ® Vario SR abutments are sterile packed incl. the enclosed abutment<br />
screw. Sterilization of the abutment as recommended before insertion<br />
is not required for immediate final integration and when the abutments<br />
remain in the CONELOG ® implant.<br />
CONELOG ®<br />
Insertion post<br />
CONELOG ®<br />
<strong>Implant</strong><br />
The angle of the Vario SR aligning tools is identical to the 20° and 30° angled<br />
CONELOG ® Vario SR abutments.<br />
80<br />
Vario SR aligning tool<br />
The angle points in the opposite<br />
direction of the implant groove.<br />
SELECTING ANGLED CONELOG ® VARIO SR ABUTMENTS<br />
WITH THE ALIGNING TOOLS 20°/30°<br />
After inserting the implant, the Vario SR 20° and 30° aligning tools can be<br />
used when multiple CONELOG ® implants are placed to check the respective<br />
insertion direction and alignment of the implant grooves before inserting<br />
the CONELOG ® Vario SR abutments. The respective aligning tool is simply<br />
placed over the insertion post attached to the implant. If necessary, the user<br />
can slightly adjust the alignment of the grooves of the inner implant configuration.<br />
The Vario SR aligning tools are packed sterile.<br />
Vario SR aligning tool 20°<br />
Vario SR aligning tool and<br />
angled CONELOG ® Vario SR<br />
abutment 20°<br />
Vario SR aligning tool 30°<br />
2 chamfer = 20° 3 chamfer = 30°<br />
Vario SR aligning tool and<br />
angled CONELOG ® Vario SR<br />
abutment 30°<br />
IMPORTANT NOTE<br />
During intraoral use, products must be secured in general against aspiration<br />
and swallowing.