CONELOG ® IMPLANT SYSTEM<br />

PROSTHETIC RESTORATIONS<br />

CONELOG ® TEMPORARY<br />

RESTORATION<br />

FABRICATION OF A TEMPORARY<br />

SINGLE-TOOTH RESTORATION<br />

EXAMPLE CHAIR-SIDE<br />

The CONELOG ® temporary abutment is inserted into the CONELOG ® implant<br />

and turned until the cams engage with the grooves of the implant.<br />

Next, the CONELOG ® abutment screw is inserted into the CONELOG ® temporary<br />

abutment and tightened by hand with a screwdriver, hex. The vestibular<br />

center and the desired occlusal height are marked on the abutment.<br />

Inserting the CONELOG ® Temporary abutment<br />

The custom shortening and/or grinding of the CONELOG ® temporary abutment<br />

is performed extraorally in order to prevent contamination of the surrounding<br />

tissue with particles from the grinding. To simplify handling, the<br />

abutment can be screwed onto a CONELOG ® lab analog or onto a CONELOG ®<br />

abutment collect for the universal holder.<br />

To protect the CONELOG ® abutment screw, we recommend using a<br />

CONELOG ® lab screw with matching diameter.<br />

CONELOG ® ART. NO.<br />

Lab screw,<br />

hex, brown anodized<br />

C4006.1601 C4006.2001<br />

<strong>Implant</strong> Ø mm 3.3/3.8/4.3 5.0<br />

Thread M 1.6 M 2.0<br />

34<br />

IMPORTANT NOTE<br />

The CONELOG ® lab screws must not be used on the patient!<br />

CONELOG ® Abutment collect with universal holder<br />

CONELOG ® Screw design