05.04.2013 Views

Introduction to Soil Chemistry

Introduction to Soil Chemistry

Introduction to Soil Chemistry

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

160 spectroscopy<br />

<strong>to</strong> scratch them. When using cuvettes that have not been used before, they<br />

should be tested <strong>to</strong> make sure that they are all the same. This is accomplished<br />

by inserting them in<strong>to</strong> the spectropho<strong>to</strong>meter and noting their absorbence.All<br />

should be the same. Keep in mind that empty cuvettes will have a higher<br />

absorbence than when filled with water. This is because light is refracted at<br />

each surface and when filled with water there is less refraction at the surfaces.<br />

8.8.1. Zeroing and Blanks<br />

When using a spectropho<strong>to</strong>meter for a colorimetric analysis, both the 0 (•A)<br />

and 100% (0A) readings must be set. Once the instrument has warmed up,<br />

with nothing in the sample compartment, the readout is set <strong>to</strong> 0. A blank, a<br />

solution containing all the components used in the analysis except the analyte<br />

being measured, is placed in a cuvette and placed in the sample compartment,<br />

and the instrument is adjusted <strong>to</strong> 100%. This procedure is intended <strong>to</strong> account<br />

for all interferences that may be introduced in<strong>to</strong> the measurement by components<br />

other than the analyte of interest. Once the instrument is adjusted,<br />

determination of the absorbence of standards and samples can be made.<br />

8.8.2. Relating Component Concentration <strong>to</strong> the Original Sample<br />

Performing an analysis requires the preparation of a standard or calibration<br />

curve. A series of standard solutions, a minimum of three, containing known<br />

amounts of the component of interest are prepared. There are two primary<br />

restrictions on these solutions: (1) they should be over the range of the<br />

expected concentrations of the component of interest and (2) all results for<br />

the extracted component must be higher than the lowest standard solution and<br />

lower than the highest standard. If a result is beyond these limits, its concentration<br />

cannot be determined. In the table of data for Figure 8.9, Unk1<br />

(unknown 1) is 0.05, which is below the lowest standard, specifically, 0.15, and<br />

so its value cannot be determined. Likewise, the Unk2 value is 0.97, which is<br />

above the highest standard and also cannot be determined.<br />

If, as with Unk 2, the value is above the highest standard, then the sample<br />

may be diluted and retested. In this case the concentration must be corrected<br />

for the amount of dilution. Similarly, in some cases when the concentration of<br />

an unknown is lower than the lowest standard, the solution can be concentrated<br />

and reanalyzed.<br />

Furthermore, standards must be over a range where there is a direct or<br />

straight-line relationship between the amount of color produced and the<br />

amount of component present. It is common for data points beyond the standard<br />

curve <strong>to</strong> be part of a different straight line or simply not measurable<br />

because the solution is <strong>to</strong>o light or <strong>to</strong>o dark. In Figure 8.9 the distance between<br />

points 0,0 and 10,0.15 may form a line very different from the one shown in<br />

the graph. We know nothing about this region. The same can be said about the<br />

region above points 90,0.95. With many instruments very low absorbances are

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!