3. Commissioning<br />
Power supply unit, Order No. 77 3401 00<br />
1. Connect input and output terminal connections. (Refer to items 2.1 and 2.2)<br />
2. Switch on the input voltage.<br />
From now on, the connected nurse call system devices are supplied with 24 V DC.<br />
The green LED "Power" is on due to the voltage at the terminal clamps.<br />
4. Spare parts<br />
Spare part Order No.<br />
Mains Fuse F1 (5 x 20) T 5,0 A 00 0130 41<br />
Output fuse F2 FKS 15 A 00 0132 02<br />
Clamp for input voltage, 3-pole 00 0211 40<br />
Clamp for 0 V and +24 V, 4-pole 00 0211 41<br />
Indicator clamp, 7-pole 00 0211 39<br />
5. <strong>Technical</strong> data<br />
Power supply unit<br />
Input<br />
Order No. 77 3401 00<br />
Nominal input voltage 115 – 230 V AC<br />
Input voltage range 90 – 264 V AC<br />
Input frequency 47 – 63 Hz<br />
Input current 4 A at 115 V AC / 2 A at 230 V AC<br />
Start-up current surge 22 A at 115 V AC / 44 A at 230 V AC, typical<br />
Power factor 0.95 at 230 V AC / 0.98 at 115 V AC under full load<br />
Efficiency<br />
Output<br />
88%, typical<br />
Nominal output current 12,5 A DC<br />
Output voltage 24 V DC +/- 2%<br />
Calculated output power 288 W<br />
Rippling of nominal output current < 100 mVeff<br />
Overload protection, power supply unit,<br />
internal<br />
< 14 A<br />
Overvoltage protection, power supply<br />
unit<br />
26.7 – 32.4 V DC, typical<br />
Overtemperature protection 75 °C +/- 5°C (power diode)<br />
output voltage switched off, automatic re-engaging<br />
EMC CE-certified EN 55022 Class B; ENV 50204<br />
EN 61000-3-2, -3; EN 61000-4-2, -3, -4, -5, -6, -8, -11<br />
Safety EN 60950<br />
Output Safety low voltage SELV<br />
Test voltage PRI – SEC 3 kV<br />
Protection class Class I<br />
Protection<br />
Safety devices<br />
IP 20<br />
Mains Fuse F1 (5 x 20) T 5.0 A<br />
Output fuse F2 FKS<br />
General<br />
15 A FKS<br />
Cooling Convection, free airflow<br />
Permissible ambient temperature 0 ... +50 °C<br />
Storage temperature -25 ... + 60 °C<br />
MTBF 100,000 hours<br />
Weight approx. 2 kg<br />
Dimensions (HxWxD) 244 x 325 x 178 mm<br />
Mounting Wall mounting<br />
Connecting clamps Screw / plug-in clamps<br />
© <strong>Tunstall</strong> <strong>GmbH</strong>, Orkotten 66, 48291 Telgte, Germany, www.tunstall.de, 00 8802 42, Rev. 3.1 (05/2008) English - 3<br />
<strong>Technical</strong> <strong>Manual</strong> Page 143